文献速递胸部技术与安全CT

韩国肺癌筛查项目中的体型特异性剂量估算:32 cm直径体模能否代表韩国标准体型患者?

Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population?
PubMed 文献 · Korean J Radiol · 2018-10-18

本研究基于韩国肺癌筛查(K-LUCAS)项目256名参与者、4家机构的数据,比较了使用32 cm直径参考体模的CTDIvol与基于体型换算因子(f-size)重新计算的SSDE,发现SSDE高于CTDIvol(2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy,p < 0.001),且SSDE/CTDIvol比值在低BMI组更高,标准体型参与者的比值高于大体型参与者且不受自动曝光控制影响。结论认为基于参考体模的CTDIvol低估了标准体型韩国参与者的低剂量CT辐射暴露。

要点

  • 基于32 cm体模的CTDIvol低估了标准体型韩国参与者的低剂量CT辐射暴露。
  • SSDE/CTDIvol比值随BMI降低而升高:肥胖1.26、超重1.37、正常1.43、偏瘦1.53。
  • 标准体型参与者的SSDE/CTDIvol比值高于大体型参与者,无论是否使用自动曝光控制。

核心结论

  • 基于参考体模的容积CT剂量指数低估了标准体型韩国参与者的低剂量CT辐射暴露。
  • 需要验证针对标准体型韩国参与者的最佳辐射剂量限值。

研究设计与样本

  • 研究类型为韩国肺癌筛查(K-LUCAS)项目预研究的剂量评估分析,参与者共256名,来自4家机构。
  • 使用32 cm直径参考体模的容积CT剂量指数(CTDIvol)与依据美国医学物理学家协会第204号报告、基于体型换算因子(f-size)重新计算的体型特异性剂量估算(SSDE)进行比较。
  • 比较按体重指数水平(偏瘦/正常 vs. 超重/肥胖)和自动曝光控制适应(是/否)进行分层评估。

诊断标准与关键数值

  • SSDE高于CTDIvol:2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy,p < 0.001,因所有参与者的f-size均大于1.0。
  • SSDE/CTDIvol比值:肥胖组(n = 103)1.26、超重组(n = 70)1.37、正常组(n = 75)1.43、偏瘦组(n = 4)1.53。
  • 使用自动曝光控制时,大体型 vs. 标准体型参与者的SSDE/CTDIvol:1.30 ± 0.08 vs. 1.44 ± 0.08(p < 0.001)。
  • 不使用自动曝光控制时,大体型 vs. 标准体型参与者的SSDE/CTDIvol:1.32 ± 0.08 vs. 1.42 ± 0.06(p < 0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea.

MATERIALS AND METHODS: For participants (n = 256, four institutions) of K-LUCAS pilot study, volume CT dose index (CTDIvol) using a 32-cm diameter reference phantom was compared with SSDE, which was recalculated from CTDIvol using size-dependent conversion factor (f-size) based on the body size, as described in the American Association of Physicists in Medicine Report 204. This comparison was subsequently assessed by body mass index (BMI) levels (underweight/normal vs. overweight/obese), and automatic exposure control (AEC) adaptation (yes/no).

RESULTS: Size-specific dose estimate was higher than CTDIvol (2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy, p < 0.001), since the f-size was larger than 1.0 for all participants. The ratio of SSDE to CTDIvol was higher in lower BMI groups; 1.26, 1.37, 1.43, and 1.53 in the obese (n = 103), overweight (n = 70), normal (n = 75), and underweight (n = 4), respectively. The ratio of SSDE to CTDIvol was greater in standard-sized participants than in large-sized participants independent of AEC adaptation; with AEC, SSDE/CTDIvol in large- vs. standard-sized participants: 1.30 ± 0.08 vs. 1.44 ± 0.08 (p < 0.001) and without AEC, 1.32 ± 0.08 vs. 1.42 ± 0.06 (p < 0.001).

CONCLUSION: Volume CT dose index based on a reference phantom underestimates radiation exposure of LDCT in standard-sized Korean participants. The optimal radiation dose limit needs to be verified for standard-sized Korean participants.

来源:Korean J Radiol,查看原文。版权归原作者所有。

引用

  1. Kim EY, Kim TJ, Goo JM, et al. Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? Korean J Radiol. 2018. PMID: 30386149. doi:10.3348/kjr.2018.19.6.1179

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。