双能CT动态与静态肺灌注血容量图像的直方图分析比较
本研究对65例患者(肺血管疾病11例、肺癌27例、肺血栓栓塞症27例)行双能CT动态与静态肺灌注血容量(LPBV)成像,比较两组图像的均值衰减、噪声、SNR、CNR及直方图偏度与峰度。静态LPBV图像均值衰减、SNR、CNR更大而峰度值更小,除左肺峰度外各参数与动态LPBV图像相关性良好,在肺血栓栓塞症患者中静态与峰值动态LPBV图像相关性尤佳。
要点
- 静态LPBV图像均值衰减、SNR、CNR高于动态LPBV图像,峰度值更低。
- 除左肺峰度外,静态与动态LPBV图像各直方图参数相关性良好。
- 肺血栓栓塞症患者中静态LPBV图像峰度与动态LPBV图像相关性好。
核心结论
- 在肺血栓栓塞症患者中,静态LPBV图像各参数与显示肺动脉优势血流的峰值动态LPBV图像相关性良好。
研究设计与样本
- 65例患者(平均年龄61.3岁,男性36例)行动态与静态LPBV检查,病种包括肺血管疾病11例、肺癌27例、肺血栓栓塞症27例;采用双能技术在肺动脉干层面先行7组动态序列扫描,再行静态LPBV扫描,动态序列中肺实质平均衰减最大的图像定义为峰值动态LPBV图像。
诊断标准与关键数值
- 静态LPBV图像较动态LPBV图像均值衰减更大(右肺24.2、左肺24.2)、SNR更高(右肺2.31、左肺2.30)、CNR更高(右肺2.40、左肺2.39)、峰度值更小(右肺1.06、左肺0.61),差异均有统计学意义(p < 0.001);除左肺峰度(r = 0.17)外,各参数与动态LPBV图像相关性良好(r = 0.4–0.77,p ≤ 0.001);静态LPBV图像峰度与动态LPBV图像峰度相关性良好(r = 0.41–0.77,p < 0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study was to compare the results of a histogram-based analysis of static and dynamic lung perfused blood volume (LPBV) images.
METHODS: Sixty-five patients (mean age: 61.3 years, 36 male) underwent dynamic and static LPBV for evaluation of pulmonary vascular diseases (n = 11), lung carcinoma (n = 27) or pulmonary thromboembolism (PTE: n = 27). Seven sets of dynamic sequential scans were performed at the pulmonary trunk using dual-energy technique before the static LPBV scan. The image of lung parenchyma that showed the greatest mean attenuation in dynamic series was defined as the peak dynamic LPBV image. The differences and correlations in the mean attenuation, image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), histogram skewness and histogram kurtosis were evaluated according to the type of disease in static and dynamic LPBV images.
RESULTS: Static LPBV images showed significantly larger mean attenuation (Rt:24.2, Lt: 24.2), SNR (Rt:2.31, Lt:2.30), and CNR (Rt:2.40, Lt:2.39), and smaller kurtosis values (Rt:1.06, Lt:0.61) values in comparison to dynamic LPBV images (p < 0.001); however, with the exception of kurtosis of the left lung (r = 0.17), these values were well-corrected with that of the dynamic LPBV images in these values (r = 0.4-0.77, p ≤ 0.001) without kurtosis of left lung (r = 0.17) in all patients. The histogram kurtosis of static LPBV image showed a good correlation with that of dynamic LPBV (r = 0.41-0.77, p < 0.05), especially in patients with PTE.
CONCLUSION: In patients with PTE, the static LPBV image valueswere well correlated with the peak dynamic LPBV images which demonstrated pulmonary artery-dominant flow.
引用