锥形束CT剂量计算技术的评估

An evaluation of techniques for dose calculation on cone beam computed tomography.
PubMed 文献 · Br J Radiol · 2019-02-26

研究评估四种锥形束CT(CBCT)剂量重算技术的准确性与效率,纳入15例接受容积旋转调强放疗的头颈、肺和前列腺患者。剂量体积直方图显示各技术与计划CT差异在0.0%至2.3%之间;变形配准在容积伽马分析中通过率最高。变形配准、HU覆盖和部位特异性校准曲线均为剂量学准确且高效的方法。

要点

  • 四种CBCT剂量重算技术与计划CT的剂量体积直方图差异在0.0%至2.3%之间。
  • 变形配准在容积伽马分析中通过率最高,50%阈值中位通过率范围因部位而异。
  • 变形配准、HU覆盖和部位特异性校准曲线均为剂量学准确且高效的方法。

核心结论

  • 变形配准、HU覆盖和部位特异性校准曲线均被认定为在CBCT图像上进行剂量计算的剂量学准确且高效的方法。

研究设计与样本

  • 研究对四种已建立的技术进行评估,并通过剂量计算评估其准确性;每种技术应用于15例接受容积旋转调强放疗的患者,治疗部位为头颈、肺和前列腺之一。

诊断标准与关键数值

  • 以计划CT剂量分布为参考,采用剂量体积直方图指标和2.0%/0.1 mm伽马分析确定计划CT与CBCT重算之间的差异;容积伽马分析中变形图像剂量重算通过率最高,50%阈值中位通过率范围:前列腺99.6-99.9%、头颈94.6-96.0%、肺94.8.0-96.0%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE:: To assess the accuracy and efficiency of four different techniques, thus determining the optimum method for recalculating dose on cone beam CT (CBCT) images acquired during radiotherapy treatments.

METHODS:: Four established techniques were investigated and their accuracy assessed via dose calculations: (1) applying a standard planning CT (pCT) calibration curve, (2) applying a CBCT site-specific calibration curve, (3) performing a density override and (4) using deformable registration. Each technique was applied to 15 patients receiving volumetric modulated arc therapy to one of three treatment sites, head and neck, lung and prostate. Differences between pCT and CBCT recalculations were determined with dose volume histogram metrics and 2.0%/0.1 mm gamma analysis using the pCT dose distribution as a reference.

RESULTS:: Dose volume histogram analysis indicated that all techniques yielded differences from expected results between 0.0 and 2.3% for both target volumes and organs at risk. With volumetric gamma analysis, the dose recalculation on deformed images yielded the highest pass-rates. The median pass-rate ranges at 50% threshold were 99.6-99.9%, 94.6-96.0%, and 94.8.0-96.0% for prostate, head and neck and lung patients, respectively.

CONCLUSION:: Deformable registration, HU override and site-specific calibration curves were all identified as dosimetrically accurate and efficient methods for dose calculation on CBCT images.

ADVANCES IN KNOWLEDGE:: With the increasing adoption of CBCT, this study provides clinical radiotherapy departments with invaluable information regarding the comparison of dose reconstruction methods, enabling a more accurate representation of a patient's treatment. It can also integrate studies in which CBCT is used in image-guided radiation therapy and for adaptive radiotherapy planning processes.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Giacometti V, King RB, Agnew CE, et al. An evaluation of techniques for dose calculation on cone beam computed tomography. Br J Radiol. 2019. PMID: 30433821. doi:10.1259/bjr.20180383

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。