双层探测器CT头颅成像:虚拟单能图像显示颅内出血与低密度脑病变的初步经验
回顾性比较58例患者双层探测器CT头颅平扫中虚拟单能图像与常规图像的诊断质量。低密度病变在白质时120 keV的对比噪声比显著增高,在灰质时40 keV更优;出血在白质时40 keV更优,在灰质时120 keV更优。主观阅读证实120 keV对灰质内出血显示最佳,40 keV额外检出白质内4个高密度病灶和灰质内1个。
要点
- 低密度病变位于白质时选高keV(120 keV),位于灰质时选低keV(40 keV)。
- 高密度出血位于灰质时选高keV(120 keV),位于白质时选低keV(40 keV)。
- 40 keV额外检出白质内4个高密度病灶和灰质内1个,提示低keV有助发现出血。
核心结论
- 双层探测器CT的虚拟单能图像可改善低密度脑实质病变和脑实质出血的显示,但效果高度依赖病变类型和位置。
研究设计与样本
- 回顾性研究,58例患者接受头颅双层探测器CT平扫,重建40至120 keV虚拟单能图像和常规图像,以ROI测量计算对比噪声比,并由两名放射科医师用5点Likert量表主观评估。
关键影像表现/征象
- 低密度病变至白质的对比噪声比在120 keV虚拟单能图像较常规图像显著增高(p ≤ 0.01)。
- 低密度病变至灰质的对比噪声比在40 keV虚拟单能图像较常规图像显著增高(p ≤ 0.0001)。
- 脑实质出血至白质的对比噪声比在40 keV虚拟单能图像较常规图像显著增高(p ≤ 0.01)。
- 脑实质出血至灰质的对比噪声比在120 keV虚拟单能图像较常规图像显著增高(p ≤ 0.01)。
- 主观阅读证实120 keV对灰质内脑实质出血的勾画最佳,两名阅片者在40 keV额外检出白质内4个高密度病灶和灰质内1个高密度病灶。
临床应用与阅片要点
- 阅片时可按病变密度和位置选择虚拟单能图像keV:低密度病变在白质用高keV,在灰质用低keV。
- 高密度出血在灰质用高keV,在白质用低keV。
- 低keV(40 keV)有助额外发现白质和灰质内的高密度出血灶。
局限性
- 原文标题与摘要注明为初步经验(Initial experience)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Retrospective comparison of diagnostic quality of virtual monoenergetic images (VMI) and conventional images (CI) reconstructed from dual-layer detector CT (DLCT) regarding intraparenchymal hemorrhage (IPH) and hypodense parenchymal lesions (HPL) of the brain.
METHODS: 58 patients underwent unenhanced DLCT of the head. CI and VMI ranging from 40 to 120 keV were reconstructed. Objective image quality was assessed using ROI-based measurements within IPH, HPL, grey matter, white matter and cerebrospinal fluid, from which contrast to noise ratio (CNR) was calculated. Two radiologists assessed IPH, HPL, artifacts and image noise on a 5-point Likert-scale. Statistical significance was determined using Wilcoxon rank sum test.
RESULTS: In comparison to conventional images, CNR of HPL to white matter was significantly increased in VMI at 120 keV (p ≤ 0.01), whereas at 40 keV, CNR to grey matter was enhanced (p ≤ 0.0001). Contrary, CNR of IPH to white matter was increased at 40 keV (p ≤ 0.01), while CNR to grey matter was improved at 120 keV (p ≤ 0.01). Subjective readings confirmed best delineation of IPH within grey matter at 120 keV. Both readers detected four additional hyperdense lesions within white and one within grey matter at 40 keV.
CONCLUSIONS: VMI obtained with DLCT can improve depiction of hypodense parenchymal lesions and intraparenchymal hemorrhage. The initial data show a great dependency on the type of pathology and on its location: hypodense lesions in white matter and hyperdense lesions in grey matter are better visualized in higher keV reconstructions, while hyperdense lesion in white matter and hypodense lesions in grey matter are better visualized at low keV values.
引用
- Lennartz S, Laukamp KR, Neuhaus V, et al. Dual-layer detector CT of the head: Initial experience in visualization of intracranial hemorrhage and hypodense brain lesions using virtual monoenergetic images. Eur J Radiol. 2018. PMID: 30396652. doi:10.1016/j.ejrad.2018.09.010