不同磁共振设备与扫描参数对体模和健康志愿者信号强度比测量的影响:对解读脑内钆信号变化的启示

The Impact of Different Magnetic Resonance Imaging Equipment and Scanning Parameters on Signal Intensity Ratio Measurements in Phantoms and Healthy Volunteers: Implications for Interpreting Gadolinium Signal Changes Within the Brain.
PubMed 文献 · Invest Radiol · 2019-01-01

本研究在Eurospin凝胶体模和15名健康志愿者脑图像上,评估不同MRI硬件与软件设置对信号强度比(SIR)的影响。结果显示,改变归一化滤波、回波时间及射频线圈单元激活等参数可产生与脑内钆对比剂高信号程度相似的SIR百分比变化,提示解读脑内钆信号变化时需考虑设备与参数差异。

要点

  • 改变归一化滤波、回波时间及射频线圈单元激活等参数可产生与脑内钆高信号相似的SIR变化。
  • 体模置于扫描仪等中心以下时SIR测量值差异达22%。
  • 健康志愿者默认序列单次扫描SIR值变异在0.64%以内。

核心结论

  • 改变常见MR采集参数和硬件可在体模和健康志愿者中复制出与脑内钆对比剂信号高信号程度相似的SIR百分比变化。

研究设计与样本

  • 在2台MRI扫描仪(3 T和1.5 T)上使用T1加权二维自旋回波序列扫描T1弛豫时间类似齿状核、脑桥、苍白球和丘脑的凝胶体模,并在15名年轻健康志愿者中重复实验。

关键影像表现/征象

  • 默认序列下,3 T多次成像时GelDN/P和GelGP/Th的SIR均值分别变化±2.20%和±0.75%,单次成像内变化更小(GelDN/P ±0.17%,GelGP/Th ±0.15%)。
  • 1.5 T时GelDN/P的SIR变化为多次成像±1.41%、单次成像±0.41%,GelGP/Th为多次成像±0.47%、单次成像±0.33%。
  • 体模置于扫描仪等中心以下时SIR测量值差异达22%。

诊断标准与关键数值

  • 体模序贯改变参数后SIR变化:归一化滤波有无为14.07%±2.43%,不同回波时间为-7.80%±0.28%,选择性激活射频线圈单元为-5.06%±0.59%。
  • 健康志愿者序贯改变参数后SIR变化:归一化滤波有无为-24.47%±2.47%,不同回波时间为-15.32%±7.71%,选择性激活射频线圈单元为-2.90%±0.78%。
  • 健康志愿者默认序列单次扫描SIR值变异在0.64%以内。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to examine the variation in signal intensity ratio (SIR) values in Eurospin gel phantoms and healthy volunteer (HV) brain images in response to different magnetic resonance imaging hardware and software settings.

MATERIALS AND METHODS: Gel phantoms with T1 relaxation times similar to the dentate nucleus (DN), pons (P), globus palladus (GP), and thalamus (Th) were scanned using a T1-weighted 2-dimensional spin-echo sequence on 2 magnetic resonance imaging scanners (3 T and 1.5 T). Imaging was performed by sequentially altering selected magnetic resonance (MR) parameters relative to a default pulse sequence, and the protocol was implemented repeatedly over 3 months. The experiment was also repeated on a cohort of 15 young HVs. Calculations of DN/P and GP/Th SIR values were derived for the images of the gels (GelDN/P and GelGP/Th) and the HVs (HVDN/P and HVGP/Th).

RESULTS: For the default sequence, the mean SIR values of GelDN/P and GelGP/Th varied by ±2.20% and ±0.75%, respectively, when measured over multiple imaging sessions (3 T). Within a single imaging session, these variations were smaller (±0.17% for GelDN/P and ±0.15% for GelGP/Th). At 1.5 T, the equivalent SIR variations for GelDN/P were ±1.41% (multiple sessions) and ±0.41% (single session), and that for GelGP/Th were ±0.47% (multiple sessions) and ±0.33% (single session).Sequential changes to the MR sequence parameters resulted in gel SIR variations as follows: 14.07% ± 2.43% (with/without normalization filters), -7.80% ± 0.28% (different echo times, TE), and -5.06% ± 0.59% (selective activation of RF coil elements). The largest variations were noted when the gels were positioned below the scanner isocenter, where SIR measurements were different by 22%.For the HVs, the SIR values were found to be consistently within 0.64% (single session) for the default sequence. Sequential changes to the MR sequence parameters resulted in SIR variations of -24.47% ± 2.47% (with/without normalization filters), -15.32% ± 7.71% (different echo times, TE), and -2.90% ± 0.78% (selective activation of RF coil elements).

CONCLUSIONS: This study has demonstrated that SIR percentage changes from baseline of a similar magnitude to brain gadolinium contrast agent signal hyperintensities can be replicated in phantom models and HVs by altering common MR acquisition parameters and hardware.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Young LK, Gandy SJ, Priba L, et al. The Impact of Different Magnetic Resonance Imaging Equipment and Scanning Parameters on Signal Intensity Ratio Measurements in Phantoms and Healthy Volunteers: Implications for Interpreting Gadolinium Signal Changes Within the Brain. Invest Radiol. 2019. PMID: 30394963. doi:10.1097/RLI.0000000000000526