脑肿瘤在1.5T与3T临床扫描仪上测得表观扩散系数可重复性的初步研究
本研究前瞻性纳入50例经组织学证实的脑肿瘤患者,同日用同一厂商1.5T和3T MR扫描仪测量肿瘤平均绝对ADC值。结果显示两场强下平均绝对ADC值无统计学显著差异,扫描仪间一致性高(kappa 0.88–0.99),提示不同场强下ADC测量可重复。
要点
- 同日同患者同肿瘤在1.5T与3T测得的平均绝对ADC值无统计学显著差异。
- 两扫描仪间一致性高,kappa系数0.88–0.99。
- 肿瘤内ADC值差异在1.5T为1.58%–4.5%,3T为1.18%–4.37%。
核心结论
- 同一患者同一天同一肿瘤在1.5T与3.0T MR系统测得的各种原发和继发脑肿瘤平均绝对ADC值无显著差异。
研究设计与样本
- 前瞻性研究,纳入50例经组织学证实的脑肿瘤患者,同日使用同一厂商两台不同场强(1.5T和3T)MR扫描仪检查。
诊断标准与关键数值
- 采用Wilcoxon配对符号秩检验比较绝对ADC值,用相关系数评估扫描仪间一致性,用单因素方差分析检验扫描仪间平均绝对ADC最高与最低值的差异,统计显著性设为p<0.05;扫描仪间一致性kappa系数0.88–0.99。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Gradient and coil systems, pulse sequence design, and imaging parameters, as well as different scanners, can influence apparent diffusion coefficient (ADC) values. The aim of this study was to evaluate the effect of two different field strengths on the reproducibility of mean absolute ADC measurements in various primary and secondary brain tumors.
METHODS: Fifty patients with histologically proven brain tumors were prospectively examined on two MR scanners from the same vendor, with different field strengths-1.5T and 3T-on the same day. Absolute ADC values were compared using the Wilcoxon matched-pairs signed-rank test. Inter-scanner agreement between two different fields in the same tumor was examined using correlation coefficients, and the discrepancy between the highest and the lowest mean absolute ADC values between scanners was tested using a one-way analysis of variance. Statistical significance was set at p < 0.05.
RESULTS: There was no statistically significant difference between mean absolute ADC values obtained on 1.5T and 3T scanners for all patients and all brain tumor types. The intratumoral difference in ADC values, averaged from two scanners in the same tumor type, ranged from 1.58 to 4.5% for 1.5T, and from 1.18 to 4.37% for 3T.Inter-scanner agreement was high, and the kappa coefficient ranged from 0.88 to 0.99, with no significant difference between obtained values on different field strengths.
CONCLUSION: Based on the results obtained in our study, there is no significant difference between mean absolute ADC values measured in various primary and secondary brain tumors at different field strengths (1.5 and 3.0T MR systems), in the same patient, and in the same tumor, measured on the same day.