文献速递神经MR

单体素1H磁共振波谱神经代谢物的昼夜稳定性与长期可重复性

Diurnal stability and long-term repeatability of neurometabolites using single voxel 1H magnetic resonance spectroscopy.
PubMed 文献 · Eur J Radiol · 2018-09-17

本研究在3 T场强下评估一维氢质子磁共振波谱(1H-MRS)的昼夜稳定性与长期可重复性,体外采用标准脑模体,体内昼夜评估纳入10名健康受试者、可重复性研究纳入6名受试者,采集后扣带回(PCG)谱线并以LCModel处理。体外代谢物昼夜变化无统计学意义,体内仅甘油磷酸胆碱(GPC,+10%,F=10.6,p=0.001)与谷氨酸+谷氨酰胺(Glx,+6%,F=5.1,p=0.018)存在统计学显著昼夜效应。多数PCG代谢物无昼夜变化,体内浓度变异低,支持在临床3 T扫描仪上使用定位MRS进行可靠的脑神经代谢物分析。

要点

  • 体外模体主要代谢物日内/日间CV分别为1-3%/2-6%,昼夜变化无统计学意义。
  • 体内仅GPC(+10%,F=10.6,p=0.001)和Glx(+6%,F=5.1,p=0.018)有显著昼夜效应,其余主要代谢物未见。
  • 体内主要代谢物ICC为0.5-0.7,SEM低(0.001-0.078),提示检测可靠性较高。

核心结论

  • 体外MRS无昼夜效应且代谢物水平变异极小,多数PCG代谢物无昼夜变化,体内代谢物浓度变异低,支持在临床3 T扫描仪上使用定位MRS进行可靠的脑神经代谢物分析。

研究设计与样本

  • 研究在3 T场强下进行体外与体内1H-MRS评估;体外使用标准脑模体,体内昼夜评估纳入10名健康受试者,可重复性研究纳入6名受试者;从后扣带回(PCG)采集MRS并以LCModel处理;昼夜效应采用重复测量方差分析,可重复性用变异系数(CV)评估,可靠性用标准误测量(SEM)与组内相关系数(ICC)评估。

关键影像表现/征象

  • 体外标准脑模体主要代谢物(NAA、Cr、mI、Glx、tCho)日内CV为1-3%、日间CV为2-6%,昼夜代谢变化无统计学意义。
  • 体内仅甘油磷酸胆碱(GPC)升高10%(F=10.6,p=0.001)和Glx升高6%(F=5.1,p=0.018)存在统计学显著昼夜效应。
  • 体内主要代谢物日内/日间受试者内CV分别为2-5%/5-9%。
  • 体内主要代谢物ICC为0.5-0.7,SEM为0.001-0.078。
  • 体内测量周间隔对代谢物比值影响极小(F=1.4,p=0.09)。

诊断标准与关键数值

  • 以LCModel处理PCG单体素1H-MRS谱线,主要代谢物包括NAA、Cr、mI、Glx与tCho。
  • 可重复性以变异系数(CV)评估,可靠性以SEM与ICC评估;主要代谢物ICC为0.5-0.7,SEM为0.001-0.078。
  • 昼夜效应以重复测量方差分析判定,体内GPC(+10%,F=10.6,p=0.001)与Glx(+6%,F=5.1,p=0.018)达统计学显著。

临床应用与阅片要点

  • 在临床3 T扫描仪上对后扣带回行定位单体素1H-MRS,主要代谢物浓度变异低,支持用于脑神经代谢物分析。
  • 判读GPC与Glx时需考虑昼夜波动(分别约+10%与+6%),其余主要PCG代谢物未见显著昼夜变化。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: This study was designed to evaluate the diurnal stability and long-term repeatability and reliability of one-dimensional (1D) hydrogen magnetic resonance spectroscopy (1H-MRS) in vitro and in vivo at 3 T.

MATERIAL AND METHOD: A standard brain phantom was used for in vitro study. In vivo diurnal evaluation involved ten healthy subjects, while repeatability study involved six subjects. MRS was acquired from posterior cingulate gyrus (PCG), and processed with LCModel. Diurnal effects were assessed with repeated measures ANOVAs, repeatability was evaluated using coefficient of variation (CV), while reliability was assessed with standard error measurement (SEM) and intra-class correlation coefficient (ICC).

RESULTS: Diurnal metabolic changes in vitro were non-significant. The intra/inter-in vitro CVs for the major metabolites; N-acetylaspartate (NAA), creatine (Cr), myo-inositol (mI), glutamate + glutamine (Glx) and total choline (tCho) were 1-3%/2-6%, respectively. Statistically significant in vivo diurnal effects were only seen for glycerophosphocholine (GPC, +10%, F = 10.6, p = 0.001) and Glx (+6%, F = 5.1, p = 0.018). The intra/inter-subject CVs for the major metabolites ranged from 2-5%/ 5-9%, respectively. The major metabolites displayed ICC ranging from 0.5-0.7 and low SEM (0.001-0.078) reflecting high reliability in detecting neurometabolites. The inter-week interval for in vivo measurements had minimal effect on metabolite ratios (F = 1.4, p = 0.09).

CONCLUSION: In vitro MRS showed no diurnal effects and minimal variation in metabolite levels. Most PCG metabolites are not altered diurnally. The low in vivo variability of metabolite concentration supports the use of localised MRS on clinical 3 T scanners for reliable neurometabolic profiling of the brain.