使用光谱探测器CT鉴别出血与碘:体模研究

Differentiation of Hemorrhage from Iodine Using Spectral Detector CT: A Phantom Study.
PubMed 文献 · AJNR Am J Neuroradiol · 2018-11-08

该体模研究将不同血细胞比容的离心血、血液-碘混合物及碘溶液置于体模中,用光谱探测器CT扫描,比较常规与虚拟平扫衰减值。虚拟平扫及相对虚拟平扫诊断出血的曲线下面积分别为0.97和0.87,联合阈值≥10 HU虚拟平扫且≥40%相对虚拟平扫的敏感度/特异度为100%/100%。

要点

  • 虚拟平扫衰减随血液成分减少而显著降低,常规CT衰减在所有含血样本间无显著差异。
  • 虚拟平扫诊断出血的AUC为0.97,相对虚拟平扫为0.87,常规CT衰减为0.29,碘定量为0.16。
  • 联合阈值≥10 HU虚拟平扫且≥40%相对虚拟平扫的敏感度/特异度为100%/100%。

核心结论

  • 光谱探测器CT在体模设置下能准确鉴别血液与碘对比剂。

研究设计与样本

  • 体模研究,使用离心血(血细胞比容5%-85%)、纯血、血液-碘混合物(75/25、50/50、25/75比例)及碘溶液(0-14 mg I/mL)在常规成像衰减12-75 HU范围内扫描,比较常规与虚拟平扫衰减。

关键影像表现/征象

  • 常规图像衰减在所有含血样本间无显著差异(P > .05)。
  • 虚拟平扫衰减随血液成分减少而显著降低(P < .01)。
  • 相对虚拟平扫值在所有研究类别间有显著差异(P < .01),且在95%置信区间内正确估计出血成分大小。

诊断标准与关键数值

  • 虚拟平扫诊断出血的AUC为0.97。
  • 相对虚拟平扫诊断出血的AUC为0.87。
  • 常规CT衰减诊断出血的AUC为0.29。
  • 碘定量诊断出血的AUC为0.16。
  • ≥10 HU虚拟平扫阈值诊断出血的敏感度/特异度为100%/23%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Conventional CT often cannot distinguish hemorrhage from iodine extravasation following reperfusion therapy for acute ischemic stroke. We investigated the potential of spectral detector CT in differentiating these lesions.

MATERIALS AND METHODS: Centrifuged blood with increasing hematocrit (5%-85%) was used to model hemorrhage. Pure blood, blood-iodine mixtures (75/25, 50/50, and 25/75 ratios), and iodine solutions (0-14 mg I/mL) were scanned in a phantom with attenuation ranging from 12 to 75 HU on conventional imaging. Conventional and virtual noncontrast attenuation was compared and investigated for correlation with calculation of relative virtual noncontrast attenuation. Values for all investigated categories were compared using the Mann-Whitney U test. Sensitivity and specificity of virtual noncontrast, relative virtual noncontrast, conventional CT attenuation, and iodine quantification for hemorrhage detection were determined with receiver operating characteristic analysis.

RESULTS: Conventional image attenuation was not significantly different among all samples containing blood (P > .05), while virtual noncontrast attenuation showed a significant decrease with a decreasing blood component (P < .01) in all blood-iodine mixtures. Relative virtual noncontrast values were significantly different among all investigated categories (P < .01), with correct hemorrhagic component size estimation for all categories within a 95% confidence interval. Areas under the curve for hemorrhage detection were 0.97, 0.87, 0.29, and 0.16 for virtual noncontrast, relative virtual noncontrast, conventional CT attenuation, and iodine quantification, respectively. A ≥10-HU virtual noncontrast, ≥20-HU virtual noncontrast, ≥40% relative virtual noncontrast, and combined ≥10-HU virtual noncontrast and ≥40% relative virtual noncontrast attenuation threshold had a sensitivity/specificity for detecting hemorrhage of 100%/23%, 89%/95%, 100%/82%, and 100%/100%, respectively.

CONCLUSIONS: Spectral detector CT can accurately differentiate blood from iodinated contrast in a phantom setting.