如何测量颅内动脉瘤尺寸:侵入性与非侵入性方法的体模研究
该体模研究将两个已知尺寸的动脉瘤模型置入人颅骨内,比较多层CT、平板血管造影CT、3D DSA和常规2D DSA等多种成像与后处理方法测量动脉瘤宽度、瘤颈和高度与已知尺寸的平均差异,发现根据先前3D扫描调整且缩短物-探测器距离的2D DSA测量最精确,CT-based图像显著劣于该标准。
要点
- 根据先前3D扫描调整且缩短物-探测器距离的2D DSA测量最精确,动脉瘤1平均差异-0.07±0.61 mm,动脉瘤2为0.17±0.39 mm。
- 常规DSA图像测量结果与最精确技术相似。
- CT-based图像测量显著劣于该标准技术。
核心结论
- 根据先前3D扫描调整且缩短物-探测器距离的2D DSA可获得最精确的动脉瘤测量,因此应在所有血管内动脉瘤治疗前进行。
研究设计与样本
- 体模研究,将两个已知尺寸的动脉瘤模型置入人颅骨内,注射碘对比剂后分别使用128层CT扫描仪或Artis Q双平面血管造影系统成像,测量动脉瘤宽度、瘤颈和高度并计算各技术与已知尺寸的平均差异,以测量最精确的技术定义为参考标准,并采用Bland-Altman图将所有技术与参考标准比较。
诊断标准与关键数值
- 根据先前3D扫描调整且缩短物-探测器距离的血管造影测量最精确:动脉瘤1平均差异-0.07±0.61 mm,动脉瘤2为0.17±0.39 mm;常规DSA图像测量结果与之相似;CT-based图像显著劣于该参考标准。
临床应用与阅片要点
- 作者结论认为,根据先前3D扫描调整且缩短物-探测器距离的2D DSA可获得最精确的动脉瘤测量,因此应在所有血管内动脉瘤治疗前进行。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Endovascular treatment of intracranial aneurysms has relevantly changed over the past decades. Multiple new devices such as intrasaccular flow diverters have broadened the treatment spectrum but require very exact aneurysm sizing. In this study, we investigated multidetector and flat panel angiographic CT and digital subtraction imaging as well as different postprocessing methods (multiplanar reconstruction, volume-rendering technique, 3D DSA, and conventional 2D angiography) for their ability to exactly size 2 aneurysm models.
MATERIALS AND METHODS: Two aneurysm models with known aneurysm sizes were placed inside a human skull. After injection of iodine contrast media, imaging was performed using a 128-slice CT scanner or an Artis Q biplane angiosuite, respectively. Aneurysms were measured for width, neck, and height, and the mean difference from the known sizes was calculated for each technique. The technique with the most exact measurement was defined as the criterion standard. We performed Bland-Altman plots comparing all techniques against the criterion standard.
RESULTS: Angiograms adjusted according a previous 3D run with a short object-to-detector distance resulted in the most exact aneurysm measurement: -0.07 ± 0.61 mm for aneurysm 1 and 0.17 ± 0.39 mm for aneurysm 2. Measurements of conventional DSA images were similar, and CT-based images were significantly inferior to the criterion standard.
CONCLUSIONS: 2D DSA with a short objective-to-detector distance adjusted according to a previous 3D run resulted in the most exact aneurysm measurement and should therefore be performed before all endovascular aneurysm treatments.