基于CT的碘浓度测量有多准确和精确?体模研究中单源与双源双能CT最小可检测浓度差的比较

How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study.
PubMed 文献 · Eur Radiol · 2018-10-01

在3D打印体模中模拟8种碘浓度(0.2-4 mgI/mL),比较四台单源与双源双能CT的碘定量绝对误差及最小可检测浓度差。患者与扫描相关因素显著影响误差和最小可检测差,现行代DECT误差低于前代,最小可检测差范围为0.4-1.5 mgI/mL。

要点

  • 体模增大和辐射剂量降低均使碘定量误差显著增高(p<0.001)。
  • 现行代双能CT的碘定量误差显著低于前代(p<0.001)。
  • 相同成像条件下最小可检测碘浓度差为0.4-1.5 mgI/mL。

核心结论

  • 患者和扫描相关因素对单源快kV切换和双源双能CT的碘浓度误差及最小可检测差均有显著影响。

研究设计与样本

  • 体模研究,在3D打印的中等和大尺寸体模中模拟8种碘浓度(0.2-4 mgI/mL),在四台不同的单源和双源双能CT上以双能模式扫描。
  • 辐射剂量:中等尺寸为7和10 mGy,大尺寸为10、13和16 mGy;碘图采用滤波反投影和厂商特定迭代重建算法重建。

诊断标准与关键数值

  • 测量碘定量的绝对误差(E),用多变量回归模型确定CT扫描仪、碘浓度、体模尺寸、辐射剂量和重建算法对E的影响。
  • 计算相同成像条件下(条件内)和不同成像条件间(条件间)碘浓度的最小可检测差(ICmin),范围为0.4-1.5 mgI/mL。
  • 现行代DECT误差显著低于前代(p<0.001);所有CT扫描仪条件下,体模尺寸增大和辐射剂量降低均使误差显著增高(p<0.001);碘浓度仅对单源DECT的误差有显著影响(p<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To assess the impact of scan- and patient-related factors on the error and the minimum detectable difference in iodine concentration among different generations of single-source (SS) fast kV-switching and dual-source (DS) dual-energy CT (DECT).

METHODS: Lesions having eight different iodine concentrations (0.2-4 mgI/mL) were emulated in a 3D-printed phantom of medium and large size. Each combination of concentration and size was scanned in dual-energy mode on four different SS and DS DECTs. Radiation doses were 7 and 10 mGy (medium size) and 10, 13, and 16 mGy (large size). Iodine maps were reconstructed with filtered back projection (FBP) and vendor-specific iterative reconstruction algorithms (IRs). Absolute error of iodine quantification (E) was measured. Multivariate regression models determined the influence of CT scanner, iodine concentration, phantom size, radiation dose, and reconstruction algorithm on E. The minimum detectable difference in iodine concentration (ICmin) under the same imaging conditions (intra-conditional) and among different imaging conditions (inter-conditional) was calculated.

RESULTS: The error was significantly lower in current than in previous DECT generations (p < 0.001). For all CT scanner conditions, the error was significantly higher with increasing phantom size and decreasing radiation dose (p < 0.001). Iodine concentration only significantly affected the error for SS DECT (p < 0.001). ICmin depended on patient- and scan-related factors and ranged from 0.4 to 1.5 mgI/mL.

CONCLUSIONS: Patient- and scan-related factors have a significant impact on the error and minimum detectable difference in iodine concentration within and among SS fast kV-switching and DS DECT.

KEY POINTS: • Patient- and scan-related factors have a significant impact on the error and minimum detectable difference in dual-energy CT-based iodine quantification. • Third-generation DECTs outperformed second-generation scanners for both single-source and dual-source dual-energy CT. • The minimum intra- and inter-conditional detectable difference in iodine concentration ranged from 0.4 to 1.5 mg iodine/mL.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Euler A, Solomon J, Mazurowski MA, et al. How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study. Eur Radiol. 2018. PMID: 30276672. doi:10.1007/s00330-018-5736-0