机械取栓中实现有效血流逆转需要球囊导引导管
在猪颈动脉模型中比较8F长鞘与8F球囊导引导管抽吸时的血流特征,发现球囊导引导管充盈球囊后均能血流停止,加抽吸后血流显著逆转;而经8F鞘抽吸仅见血流减少、血管塌陷或振荡血流,净血流可前向或后向。
要点
- 球囊导引导管充盈球囊即可可靠实现血流停止,加抽吸后血流逆转显著。
- 经8F鞘抽吸即使泵抽吸仍有前向血流,手动抽吸可出现血管塌陷或振荡血流。
- 经鞘抽吸时净血流方向不定,可前向或后向,不能可靠实现血流逆转。
核心结论
- 仅球囊导引导管联合注射器或高流量泵抽吸能可靠实现血流停止和血流逆转。
- 经8F鞘抽吸会导致血管塌陷或振荡血流,净血流可前向或后向。
研究设计与样本
- 研究类型为猪体内模型实验,比较8F长鞘与8F球囊导引导管的颈总动脉血流特征。
- 使用VacLok注射器手动抽吸和Penumbra泵(Pump MAX)抽吸,通过连续双功能超声评估血流动力学。
关键影像表现/征象
- 8F鞘泵抽吸后血流显著减少(225 ± 25 mL/min,P < .001),但仍存在前向血流。
- 8F鞘手动抽吸7次测量中2次血管塌陷,其余5次呈振荡血流(收缩期前向、舒张期后向)。
- 8F鞘手动抽吸后净血流3次前向(52 ± 44 mL/min)、2次后向(-95 ± 52 mL/min)。
- 球囊导引导管充盈球囊均导致血流停止。
- 球囊导引导管加泵抽吸或手动抽吸后血流显著逆转,分别为-1100 ± 230 mL/min和-468 ± 46 mL/min(均P < .001)。
临床应用与阅片要点
- 球囊导引导管可可靠实现血流停止和逆转,经8F鞘抽吸不能可靠实现血流逆转。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Blood flow management in the carotid artery during mechanical thrombectomy is crucial for safety and effectiveness. There is an ongoing discussion about whether balloon-guide catheters or large-bore sheaths are needed for effective flow management. We compared general flow characteristics of proximal aspiration through a large-bore sheath and a balloon-guide catheter in a porcine in vivo model.
MATERIALS AND METHODS: We investigated blood flow in a porcine common carotid artery with and without aspiration (VacLok syringe and Penumbra pump, Pump MAX) through an 8F-long sheath and an 8F balloon-guide catheter. Blood hemodynamics were assessed via continuous duplex sonography.
RESULTS: Average vessel diameter and baseline blood flow were 4.4 ± 0.2 mm and 244 ± 20 mL/min, respectively. For the 8F sheath, pump aspiration resulted in a significant flow reduction (225 ± 25 mL/min, P < .001), but with a persisting antegrade stream. Manual aspiration resulted in collapse of the vessel in 2 of 7 measurements and oscillatory flow with antegrade systolic and retrograde diastolic components in the remaining 5 measurements. Net flow was antegrade (52 ± 44 mL/min) in 3 and retrograde (-95 ± 52 mL/min) in the remaining 2 measurements. For balloon-guide catheters, balloon inflation always resulted in flow arrest. Additional pump or manual aspiration resulted in significant flow reversal of -1100 ± 230 and -468 ± 46 mL/min, respectively (both, P < .001).
CONCLUSIONS: Only balloon-guide catheters allow reliable blood flow arrest and flow reversal in combination with aspiration via syringes or high-flow pump systems. Aspiration through an 8F sheath results in either collapse of the vessel or oscillatory flow, which can result in a net antegrade or retrograde stream.