基于低剂量双平面X线摄影的三维后足对线测量:一项临床可行性研究

3D hindfoot alignment measurements based on low-dose biplanar radiographs: a clinical feasibility study.
PubMed 文献 · Skeletal Radiol · 2018-10-23

这项前瞻性研究对50例患者比较了基于低剂量双平面X线摄影(BPR)的三维后足对线(HA)测量与长轴位X线片(LAR)二维HA测量。两位肌骨放射科医师独立测量,BPR的测量者间一致性极佳(ICC=0.992),LAR为ICC=0.962。作者建议日常评估HA时使用三维BPR测量,因为二维LAR单独使用可能在个别患者中出现较大测量误差。

要点

  • 三维BPR后足对线测量的测量者间一致性极佳,ICC=0.992(95%CI 0.986-0.995)。
  • 二维LAR测量的测量者间一致性同样极佳,ICC=0.962(95%CI 0.932-0.978),但个别患者误差较大。
  • 两种方法平均差值为-2.43°(读者1)和-2.6°(读者2),Bland-Altman图上有少数测量值落在±1.96 SD区间外。

核心结论

  • 三维BPR后足对线测量在临床环境中具有极佳的测量者间一致性,单独使用二维LAR可能在个别患者中产生较大测量误差,作者建议日常评估后足对线时使用三维BPR测量。

研究设计与样本

  • 前瞻性研究,经当地机构审查委员会批准,纳入50例患者(29例女性,平均年龄47±16.6岁),比较三维BPR与二维LAR的后足对线测量,由两位独立肌骨放射科医师(读者1和读者2)在两次独立读片 session 中完成测量,三维BPR测量使用自制MATLAB代码。

临床应用与阅片要点

  • 作者建议日常评估后足对线时使用三维BPR测量,该测量不受足部旋转位置不良影响。

局限性

  • Bland-Altman图上读者1有3个测量值、读者2有6个测量值落在±1.96 SD区间外,提示个别患者中可能出现较大测量误差。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To test a 3D-hindfoot alignment (HA) measurement technique based on low-dose biplanar radiographs (BPRs) in a clinical setting and compare the results with 2D-HA measurements on long axial view radiographs (LARs).

MATERIALS AND METHODS: This prospective study was approved by the local institutional review board. HA measurements on 3D-BPR and 2D-LAR of 50 patients (29 female; mean age 47 ± 16.6 years) were compared (positive values = valgus; negative values = varus). Two independent musculoskeletal radiologists (readers 1 and 2) performed 3D-HA measurements on BPR using a custom-made MATLAB code and measured HA on LAR during two separate readout sessions. Descriptive statistics and intraclass correlation coefficients (ICC) were calculated, and Bland-Altman plots were used for intermethod comparison.

RESULTS: Using BPRs, HA was 0.8° ± 9.°1 (range, -20.2 to 20.0) for reader 1, and 0.7° ± 9.5° (range, -21.2 to 18.3) for reader 2. HA on LARs was -2.0 ° ± 7.0° (range, -27.0° to 11.1°) for reader 1 and - 1.7° ± 7.0° (range, -24.1° to 14.3°) for reader 2. Interreader agreement for measurements was excellent, both for BPRs (ICC = 0.992; 95% CI:0.986-0.995) and LAR measurements (ICC = 0.962; 95% CI:0.932-0.978). Mean difference between the two methods was -2.43° (range, -29.4° to 25.6°) for reader 1 and -2.6° (range,-28.7° to 30.2°) for reader 2. On Bland-Altman plots, three measurements of reader 1 and six measurements of reader 2 were outside of the ±1.96 SD interval.

CONCLUSION: Hindfoot alignment measurements on 3D-BPR have an excellent interreader agreement in a clinical setting. Large measurement errors can occur in individual patients using 2D-LAR alone. Therefore, we suggest using 3D-BPR measurements in daily routine for the assessment of HA, which are independent of rotational foot malpositioning.

来源:Skeletal Radiol,查看原文。版权归原作者所有。