钆塞酸增强各向同性高分辨率三维磁共振胆管成像联合迭代去噪算法评估活体肝移植供肝胆道解剖的临床可行性
回顾性纳入75例活体肝移植供肝者,在3 T扫描仪上比较传统三维多层面T2W-MRC与钆塞酸增强各向同性高分辨率三维T1W-MRC(联合迭代去噪算法)评估胆道解剖的价值,以术中实时荧光胆管造影为参考标准。结果显示T1W-MRC采集时间更短、图像质量更优,对胆管变异和右半肝切除术中预期胆管开口数的诊断一致性更高。
要点
- IHR-3D-T1W-MRC平均采集时间17秒,显著短于3D-T2W-MRC的367秒(P < 0.001)。
- 与3D-T2W-MRC相比,IHR-T1W-MRC对肝内胆管可见度和锐利度均显著改善(所有P < 0.05),整体图像质量更高(P < 0.01)。
- IHR-T1W-MRC对胆管变异的符合率显著高于3D-T2W-MRC(87.6% vs 81.3%,P = 0.03),对预期胆管开口数的符合率也显著更高(76.9% vs 70.2%,P = 0.006)。
核心结论
- 联合迭代去噪算法的各向同性高分辨率T1W-MRC较3D-T2W-MRC显著改善胆管图像质量,并在右半肝供肝切除术中更准确地显示胆道解剖和胆管开口。
研究设计与样本
- 经机构审查委员会批准的回顾性研究,共纳入75例活体肝移植供肝者,均在3 T扫描仪上接受传统3D多层面T2W-MRC和IHR-3D-T1W-MRC检查并后续行右半肝切除术供肝。
关键影像表现/征象
- IHR-3D-T1W-MRC采用3D VIBE Dixon序列联合迭代去噪算法获得轴位和冠状位图像,迭代去噪算法基于小波阈值处理噪声水平、g因子和k空间滤波的三维复数数据。
- 与3D-T2W-MRC相比,IHR-T1W-MRC在所有评估的肝内胆管中可见度和锐利度均显著改善(所有P < 0.05)。
- IHR-T1W-MRC的整体图像质量显著高于3D-T2W-MRC(P < 0.01)。
- IHR-T1W-MRC对右肝管长度估计的观察者间一致性为优(组内相关系数0.94),3D-T2W-MRC为良(组内相关系数0.78)。
诊断标准与关键数值
- 以术中实时荧光胆管造影由外科医生共识确定的胆管变异为参考标准,IHR-T1W-MRC对胆管变异的符合率为87.6%,3D-T2W-MRC为81.3%(P = 0.03)。
- 以术中实时荧光胆管造影为参考标准,IHR-T1W-MRC对预期胆管开口数的符合率为76.9%,3D-T2W-MRC为70.2%(P = 0.006)。
- 3位委员会认证放射科医师独立评分,对胆管可见度、锐利度和整体图像质量采用5分制评分。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to evaluate the clinical feasibility of gadoxetic acid-enhanced isotropic high-resolution (IHR) 3-dimensional (3D) T1-weighted (T1W) magnetic resonance cholangiography (MRC) using an iterative denoising (ID) algorithm for evaluation of the biliary anatomy of living liver donors in comparison with conventional 3D multislice T2-weighted (T2W) MRC.
MATERIALS AND METHODS: In this institutional review board-approved retrospective study, a total of 75 living liver donors who underwent conventional 3D multislice T2W-MRC and IHR-3D-T1W-MRC on a 3 T scanner and subsequent right hepatectomy for liver donation were included. Isotropic high-resolution T1W-MRCs were obtained in both axial and coronal planes using the 3D VIBE Dixon sequence and an ID algorithm implemented with wavelet thresholding of 3D complex-valued data of the noise level, g-factor, and k-space filtering. Thereafter, 3 board-certified radiologists independently reviewed the examinations for visibility and sharpness of the bile ducts (BDs), as well as overall image quality on a 5-point scale. For diagnostic performance, anatomic variations of the BD, length of right hepatic duct, and the expected number of BD openings at right hepatectomy were also recorded. As the reference standard, BD variation was determined by surgeons in consensus using intraoperative real-time fluorescent cholangiography.
RESULTS: Mean acquisition times of 3D-T2W-MRC and IHR-T1W-MRC were 367 seconds and 17 seconds (P < 0.001), respectively. Compared with 3D-T2W-MRCs, IHR-T1W-MRCs yielded significantly improved visibility and sharpness of all evaluated intrahepatic bile ducts (all Ps < 0.05), and higher overall image quality (P < 0.01). The IHR-T1W-MRCs also demonstrated significantly higher agreement in BD variation (87.6% vs 81.3%, P = 0.03) and expected BD openings (76.9% vs 70.2%, P = 0.006) than 3D-T2W-MRC compared with the reference standard. Interobserver agreement in estimating the length of right hepatic duct, IHR-T1W-MRC showed excellent interobserver agreement (intraclass correlation coefficient, 0.94), whereas 3D-T2W-MRC showed good interobserver agreement (intraclass correlation coefficient, 0.78).
CONCLUSIONS: Isotropic high-resolution T1W-MRCs with ID provided significantly improved BD image quality and more accurate depiction of the BD anatomy and BD openings at right donor hemihepatetomy than 3D-T2W-MRC.
引用
- Kang HJ, Lee JM, Ahn SJ, et al. Clinical Feasibility of Gadoxetic Acid-Enhanced Isotropic High-Resolution 3-Dimensional Magnetic Resonance Cholangiography Using an Iterative Denoising Algorithm for Evaluation of the Biliary Anatomy of Living Liver Donors. Invest Radiol. 2019. PMID: 30281556. doi:10.1097/RLI.0000000000000512
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。