自动瓣膜追踪用于四维血流MRI流量定量

Automated Cardiac Valve Tracking for Flow Quantification with Four-dimensional Flow MRI.
PubMed 文献 · Radiology · 2018-10-30

本回顾性研究纳入114例获得性或先天性心脏病患者和46名健康志愿者,比较四维血流MRI自动瓣膜追踪与手动瓣膜追踪。自动追踪分析时间更短,四瓣膜净前向容积变异更小,反流分数评估的观察者内和观察者间一致性良好至优秀;但部分肺动脉瓣、二尖瓣和三尖瓣的反流严重程度判定不一致。

要点

  • 自动瓣膜追踪较手动追踪显著缩短四维血流MRI分析时间(14分钟 vs 25分钟,P < .001)。
  • 自动追踪四瓣膜净前向容积变异小于手动追踪(4.9% vs 9.8%,P < .001)。
  • 自动追踪净前向容积的观察者内和观察者间ICC均≥0.99,反流分数观察者内ICC≥0.94、观察者间ICC≥0.89。

核心结论

  • 在获得性或先天性心脏病患者及健康志愿者中,自动瓣膜追踪可缩短四维血流MRI分析时间并提高瓣膜流量定量的可靠性。

研究设计与样本

  • 回顾性研究,纳入114例患者(33例获得性、81例先天性心脏病;中位年龄17岁,IQR 13-49岁;51例女性、63例男性)和46名志愿者(中位年龄28岁,IQR 22-36岁;19例女性、27例男性),2006年至2017年行1.5 T或3.0 T四维血流MRI;采用每个瓣膜两个正交 cine 视图进行瓣膜追踪,以Wilcoxon符号秩检验比较分析时间、净前向容积和反流分数,以组内相关系数(ICC)检验观察者内和观察者间变异。

诊断标准与关键数值

  • 自动追踪与手动追踪分析时间:全部患者14分钟(IQR 12-15分钟)vs 25分钟(IQR 20-25分钟),P < .001;全部患者四瓣膜净前向容积变异:自动追踪4.9%(IQR 3.3%-6.7%)vs 手动追踪9.8%(IQR 5.1%-14.7%),P < .001;自动追踪净前向容积观察者内和观察者间ICC均≥0.99,反流分数观察者内ICC≥0.94、观察者间ICC≥0.89;反流严重程度判定不一致的瓣膜包括7个肺动脉瓣、22个二尖瓣和21个三尖瓣。

临床应用与阅片要点

  • 自动瓣膜追踪可用于四维血流MRI的瓣膜流量定量,缩短分析时间并提高净前向容积和反流分数评估的一致性;但肺动脉瓣、二尖瓣和三尖瓣的反流严重程度判定在两种方法间存在不一致,阅片和报告时需注意。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To compare four-dimensional flow MRI with automated valve tracking to manual valve tracking in patients with acquired or congenital heart disease and healthy volunteers. Materials and Methods In this retrospective study, data were collected from 114 patients and 46 volunteers who underwent four-dimensional flow MRI at 1.5 T or 3.0 T from 2006 through 2017. Among the 114 patients, 33 had acquired and 81 had congenital heart disease (median age, 17 years; interquartile range [IQR], 13-49 years), 51 (45%) were women, and 63 (55%) were men. Among the 46 volunteers (median age, 28 years; IQR, 22-36 years), there were 19 (41%) women and 27 (59%) men. Two orthogonal cine views of each valve were used for valve tracking. Wilcoxon signed-rank test was used to compare analysis times, net forward volumes (NFVs), and regurgitant fractions. Intra- and interobserver variability was tested by using intraclass correlation coefficients (ICCs). Results Analysis time was shorter for automated versus manual tracking (all patients, 14 minutes [IQR, 12-15 minutes] vs 25 minutes [IQR, 20-25 minutes]; P < .001). Although overall differences in NFV and regurgitant fraction were comparable between both methods, NFV variation over four valves was smaller for automated versus manual tracking (all patients, 4.9% [IQR, 3.3%-6.7%] vs 9.8% [IQR, 5.1%-14.7%], respectively; P < .001). Regurgitation severity was discordant for seven pulmonary valves, 22 mitral valves, and 21 tricuspid valves. Intra- and interobserver agreement for automated tracking was excellent for NFV assessment (intra- and interobserver, ICC ≥ 0.99) and strong to excellent for regurgitant fraction assessment (intraobserver, ICC ≥ 0.94; interobserver, ICC ≥ 0.89). Conclusion Automated valve tracking reduces analysis time and improves reliability of valvular flow quantification with four-dimensional flow MRI in patients with acquired or congenital heart disease and in healthy volunteers. © RSNA, 2018 Online supplemental material is available for this article. See also the editorial by François in this issue.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Kamphuis VP, Roest AAW, Ajmone Marsan N, et al. Automated Cardiac Valve Tracking for Flow Quantification with Four-dimensional Flow MRI. Radiology. 2018. PMID: 30375924. doi:10.1148/radiol.2018180807

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。