文献速递胸部技术与安全DR

管电压与骨抑制技术胸片上肺结节物理图像质量的关系

Relationship Between Tube Voltage and Physical Image Quality of Pulmonary Nodules on Chest Radiographs Obtained Using the Bone-Suppression Technique.
PubMed 文献 · Acad Radiol · 2018-09-28

研究使用 anthropomorphic 胸部模体(原文为 anthropomorphic chest phantom)和直径12毫米、约+100亨氏单位的球形模拟结节,在四个管电压(70、90、110、130 kVp)下获取120张胸片并生成骨抑制图像,分析对比度和对比噪声比(CNR)。结果显示在与肋骨重叠的肺野,较低管电压下对比度和CNR倾向于更高;结论认为对相对致密结节,骨抑制胸片在与肋骨重叠区域降低管电压可显著改善对比度和CNR。

要点

  • 骨抑制胸片上,与肋骨重叠区域的肺结节在较低管电压下对比度和CNR倾向于更高。
  • 不与肋骨重叠区域,仅70 kVp与90 kVp间对比度差异有统计学意义(p=0.01)。
  • 与一根或两根肋骨重叠区域,70 kVp与其他三个管电压的对比度比较p值均小于0.01。

核心结论

  • 对相对致密结节,骨抑制胸片在与肋骨重叠的肺野降低管电压可显著改善对比度和CNR。

研究设计与样本

  • 模体研究,使用 anthropomorphic 胸部模体和直径12毫米、约+100亨氏单位的球形模拟结节,肺野按与肋骨重叠情况分三区,每区定义10个结节位置,共30个位置;用70、90、110、130 kVp四个管电压获取120张胸片并生成骨抑制图像。

诊断标准与关键数值

  • 所有四个管电压两两比较采用双侧Wilcoxon符号秩检验分析对比度和CNR差异;不与肋骨重叠区仅70 kVp与90 kVp间对比度差异有统计学意义(p=0.01);与一根或两根肋骨重叠区,70 kVp与其他三个管电压的对比度比较p值均小于0.01。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Image quality of chest radiographs is affected by tube voltage. This study aimed to clarify the relationship between tube voltage and physical image quality of pulmonary nodules on bone-suppressed chest radiographs.

MATERIALS AND METHODS: An anthropomorphic chest phantom and a spherical simulated nodule, with a 12-mm diameter and approximately +100 Hounsfield units were used. The lung field of the phantom was divided into three areas, based on the overlap with the ribs in the chest radiograph. Ten positions of the simulated nodule were defined in each area. One hundred and twenty chest radiographs were acquired using four tube voltages (70 kVp, 90 kVp, 110 kVp, and 130 kVp) for a total of 30 nodule positions and were processed to create bone-suppressed images. Differences in contrast and contrast-to-noise ratio (CNR) were analyzed for all pairs of the four tube voltages using a two-sided Wilcoxon signed-rank test.

RESULTS: In the area not overlapping with ribs, a statistically significant difference was observed only in contrast between tube voltage of 70 kVp and 90 kVp (p = 0.01). In the area overlapping with one or two ribs, the contrast and CNR tended to be higher at a lower tube voltage. In particular, the p values between the contrast at 70 kVp and that at the other three tube voltage settings were less than 0.01.

CONCLUSION: For a relatively dense nodule, the contrast and CNR in the bone-suppressed chest radiograph were significantly improved with lower tube voltage in the lung field overlapping with the ribs.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Takagi S, Yaegashi T, Ishikawa M, et al. Relationship Between Tube Voltage and Physical Image Quality of Pulmonary Nodules on Chest Radiographs Obtained Using the Bone-Suppression Technique. Acad Radiol. 2018. PMID: 30269955. doi:10.1016/j.acra.2018.08.017

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。