探索配备自适应准直器的临床质子束用于临床前研究的可行性
本研究用TOPAS蒙特卡罗工具包对Mevion S250i HYPERSCAN临床质子系统的能量调制系统与自适应准直器建模,在小鼠肺癌CT图像上设计7种质子束方案,均可给予肿瘤2 Gy且覆盖良好(D95>1.95 Gy、D5<2.3 Gy),对危及器官损伤可能小于光子计划,治疗时间低于5 s。
要点
- 自适应准直器可形成小至5×1 mm²的照射野,适合小动物毫米级束流照射。
- 7种质子方案均实现肿瘤覆盖D95>1.95 Gy且D5<2.3 Gy。
- 极小野低能量时到达靶区的质子数降至1–3%,但治疗时间仍低于5 s。
核心结论
- 配备自适应准直器的临床质子束所设计的方案可用于动物照射,作者认为该研究首次之一证明其临床前研究可行性。
研究设计与样本
- 研究为质子束流建模与计划设计研究,使用TOPAS蒙特卡罗模拟工具包对Mevion S250i HYPERSCAN临床质子系统的能量调制系统与自适应准直器建模。
- 在小鼠肺癌CT图像上进行蒙特卡罗剂量计算,比较7种质子方案与精确光子照射计划,并评估治疗时间。
关键影像表现/征象
- 自适应准直器允许小至5×1 mm²的照射野用于照射,能量调制系统使230 MeV束流名义射程可按2.1 mm的倍数递减。
- 7种质子方案对肿瘤覆盖良好(D95>1.95 Gy,D5<2.3 Gy),对危及器官的损伤可能小于光子计划。
- 极小野和低能量条件下到达靶区的质子数降至1–3%,但治疗时间低于5 s。
诊断标准与关键数值
- 质子方案以给予肿瘤2 Gy为目标,从能量调制系统允许的离散能量、不同照射野大小和束流方向中优选得出。
- 剂量覆盖标准为D95>1.95 Gy且D5<2.3 Gy。
临床应用与阅片要点
- 作者结论认为,本研究中经自适应准直器准直的质子方案可用于动物照射。
- 小野低能量时质子数明显减少至1–3%,原文未据此给出临床操作建议。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE:: To investigate whether the Mevion S250i with HYPERSCAN clinical proton system could be used for pre-clinical research with millimetric beams.
METHODS:: The nozzle of the proton beam line, consisting of an energy modulation system (EMS) and an adaptive aperture (AA), was modelled with the TOPAS Monte Carlo Simulation Toolkit. With the EMS, the 230 MeV beam nominal range can be decreased in multiples of 2.1 mm. Monte Carlo dose calculations were performed in a mouse lung tumour CT image. The AA allows fields as small as 5 × 1 mm2 to be used for irradiation. The best plans to give 2 Gy to the tumour were derived from a set of discrete energies allowed by the EMS, different field sizes and beam directions. The final proton plans were compared to a precision photon irradiation plan. Treatment times were also assessed.
RESULTS:: Seven different proton beam plans were investigated, with a good coverage to the tumour (D95 > 1.95 Gy, D5 < 2.3 Gy) and with potentially less damage to the organs at risk than the photon plan. For very small fields and low energies, the number of protons arriving to the target drops to 1-3%, nevertheless the treatment times would be below 5 s.
CONCLUSION:: The proton plans made in this study, collimated by an AA, could be used for animal irradiation.
ADVANCES IN KNOWLEDGE:: This is one of the first study to demonstrate the feasibility of pre-clinical research with a clinical proton beam with an adaptive aperture used to create small fields.
引用
- Almeida IP, Vaniqui A, Schyns LE, et al. Exploring the feasibility of a clinical proton beam with an adaptive aperture for pre-clinical research. Br J Radiol. 2018. PMID: 30362812. doi:10.1259/bjr.20180446
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。