头颈部影像中双源与单源双能CT的比较
该前瞻性随机研究将102例经组织学证实为头颈癌的患者随机分配至辐射剂量匹配的单源双能CT(SS-DECT,n=51)或第三代双源双能CT(DS-DECT,n=51)检查组,重建常规图像及40和60 keV虚拟单能图像。结果显示两种技术图像质量均具诊断性,DS-DECT在血管和肿瘤衰减方面显著优于SS-DECT,40 keV和60 keV重建的主观图像质量在DS-DECT组显著更好。作者认为联合常规图像与低keV重建可获得最高整体图像质量。
要点
- 两种双能CT技术图像质量均具诊断性,但双源技术结果显著更好。
- 联合常规图像与低keV重建可获得头颈部最高整体图像质量。
- DS-DECT在血管和肿瘤衰减及主观图像质量上显著优于SS-DECT。
核心结论
- 在分滤过单源双能CT和双源双能CT中,联合常规图像与低keV重建可获得头颈部影像最高整体图像质量;DS-DECT在主观图像质量和血管及肿瘤衰减方面优于SS-DECT。
研究设计与样本
- 前瞻性随机研究纳入102例经组织学证实为头颈癌的患者,随机分配至辐射剂量匹配的SS-DECT组(n=51,120 kV,分滤过技术,384参考mAs)或DS-DECT组(n=51,80/Sn150 kV,管A 100/管B 67参考mAs),重建常规图像及40和60 keV虚拟单能图像,客观图像质量以剂量归一化对比噪声比(CNRD)评价,主观图像质量采用5点Likert量表评分。
关键影像表现/征象
- 两组中血管和肿瘤衰减的最高CNRD值均在40 keV获得。
- DS-DECT在血管和肿瘤衰减方面显著优于SS-DECT。
- SS-DECT组整体主观图像质量以常规图像最高,其次为40 keV和60 keV。
- DS-DECT组整体主观图像质量以40 keV最高,其次为60 keV和常规图像。
- 40 keV和60 keV的主观图像质量在DS-DECT组显著优于SS-DECT组(均p<0.01)。
诊断标准与关键数值
- 主观图像质量采用5点Likert量表评分。
- 客观图像质量以剂量归一化对比噪声比(CNRD)评价。
- 40 keV和60 keV的主观图像质量在DS-DECT组显著优于SS-DECT组(均p<0.01)。
临床应用与阅片要点
- 作者认为联合常规图像与低keV重建可获得最高整体图像质量。
- 作者认为本研究结果可能对CT扫描仪更换决策及考虑图像质量和后续治疗决策的针对性患者检查具有相关性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to compare image quality of single-source dual-energy CT (SS-DECT) with third-generation dual-source dual-energy CT (DS-DECT) in head and neck cancer.
MATERIALS AND METHODS: One hundred two patients with histologically proven head and neck cancer were prospectively randomized to undergo radiation dose-matched SS-DECT (n = 51, 120 kV, split-filter technique, 384 ref. mAs) or DS-DECT (n = 51, 80/Sn150 kV, tube A 100/tube B 67 ref. mAs). Inline default images (DI) and virtual monoenergetic images (VMI) for two different low energies (40 and 60 keV) were reconstructed. Objective image quality was evaluated as dose-normalized contrast to noise ratio (CNRD), and subjective image quality was rated on a 5-point Likert scale.
RESULTS: In both groups, highest CNRD values for vessel and tumor attenuation were obtained at 40 keV. DS-DECT was significantly better than SS-DECT regarding vessel and tumor attenuation. Overall subjective image quality in the SS-DECT group was highest on the DI followed by 40 keV and 60 keV. In the DS-DECT group, subjective image quality was highest at 40 keV followed by 60 keV and the DI. Forty kiloelectron volts and 60 keV were significantly better in the DS-DECT compared to the SS-DECT group (both p < 0.01).
CONCLUSIONS: In split-filter SS-DECT as well as in DS-DECT, highest overall image quality in head and neck imaging can be obtained with a combination of DI and low keV reconstructions. DS-DECT is superior to split-filter SS-DECT in terms of subjective image quality and vessel and tumor attenuation.
KEY POINTS: • Image quality was diagnostic with both dual-energy techniques; however, the dual-source technique delivered significantly better results. • Highest overall image quality in head and neck imaging can be obtained with a combination of default images and low keV reconstructions with both dual-energy techniques. • The results of this study may have relevance for the decision-making process regarding replacement of CT scanners and focused patient examination considering image quality and subsequent therapeutic decision-making.
引用
- May MS, Wiesmueller M, Heiss R, et al. Comparison of dual- and single-source dual-energy CT in head and neck imaging. Eur Radiol. 2018. PMID: 30338365. doi:10.1007/s00330-018-5762-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。