量化立体定向小鼠微图像引导放射治疗系统的摆位不确定性
研究用立体定向固定架固定小鼠颅骨,以锥形束CT(CBCT)评估小鼠间及小鼠内重复摆位的准确性。小鼠间和小鼠内立体定向摆位与CBCT对准的平均矢量偏差分别为0.78±0.27 mm和0.82±0.34 mm。作者认为,允许毫米级位置不确定性的颅脑照射可仅依赖单次每日CBCT获得的立体定向坐标;亚毫米精度需每只小鼠每日图像引导。
要点
- 小鼠间与小鼠内立体定向摆位相对CBCT对准的平均矢量偏差分别为0.78±0.27 mm与0.82±0.34 mm。
- 允许毫米级位置不确定性的颅脑照射可仅依赖单次每日CBCT获得的立体定向坐标。
- 亚毫米精度照射需对每只小鼠进行每日图像引导。
核心结论
- 允许毫米级位置不确定性的颅脑照射可仅依赖单次每日CBCT获得的立体定向坐标;需亚毫米精度的亚区域照射则需每只小鼠每日图像引导。
研究设计与样本
- 使用X-RAD SmART辐照器的临床前研究:小鼠间重复性采用无胸腺裸鼠(N=5,×4组),小鼠内重复性采用C57BL/6小鼠(N=4,每日成像共7天),均以CBCT为参考标准。
诊断标准与关键数值
- 小鼠间分析:立体定向摆位与CBCT对准的平均矢量偏差为0.78±0.27 mm(N=60次测量)。
- 小鼠内分析:立体定向摆位与CBCT对准的平均矢量偏差为0.82±0.34 mm(N=24次测量)。
临床应用与阅片要点
- 允许毫米级位置不确定性的颅脑照射可仅依赖单次每日CBCT获得的立体定向坐标。
- 亚毫米精度照射需对每只小鼠进行每日图像引导。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE:: Investigate the reproducibility of murine cranial positioning using solely a stereotactic stage, and quantify the potential improvements from the on-board image guidance of the X-RAD SmART irradiator.
METHODS:: For intermouse reproducibility, athymic nude mice (N = 5, ×4 groups) were cranially fixed on a stereotactic stage. Each mouse was imaged via cone-beam CT (CBCT). A virtual isocenter target was placed in the brain, the stage shifted to that target, and the couch positions recorded. The mouse was removed from the stage and this process repeated twice (N=60 measurements). The first acquired CBCT coordinates (within each group of five mice) were used to define "stereotactic couch coordinates." CBCT shifts were calculated to quantify the accuracy of setup based on couch coordinates alone. For intramouse reproducibility, C57BL/6 mice (N=4) were imaged daily for 7 days. Each mouse had individual stereotactic coordinates defined from their first day of CBCT localization, and positional shifts required on the six subsequent days of imaging were quantified (N = 24 measurements).
RESULTS:: The mean vector shift between stereotactic setup and CBCT alignment for inter and intramouse analysis was 0.78 ± 0.27 mm and 0.82 ± 0.34 mm, respectively.
CONCLUSION:: Cranial irradiation that can permit positional uncertainties on the order of a millimeter can rely solely on stereotactic coordinates derived from a single daily CBCT. Irradiations of subregions requiring submillimeter accuracy require daily image guidance for each mouse.
ADVANCES IN KNOWLEDGE:: This is the first investigation of stereotactic reproducibility using the X-RAD SmART and it suggests a method for increased efficiency in high-throughput experiments.
引用
- Walb MC, Carlson BL, Sarkaria JN, et al. Quantifying the setup uncertainty of a stereotactic murine micro-image guided radiation therapy system. Br J Radiol. 2018. PMID: 30299986. doi:10.1259/bjr.20180487
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。