多动态多回波序列在3 T下快速同步弛豫测量的线性、偏倚、扫描仪内重复性与扫描仪间再现性:标准化体模与健康对照验证研究

Linearity, Bias, Intrascanner Repeatability, and Interscanner Reproducibility of Quantitative Multidynamic Multiecho Sequence for Rapid Simultaneous Relaxometry at 3 T: A Validation Study With a Standardized Phantom and Healthy Controls.
PubMed 文献 · Invest Radiol · 2019-01-01

本研究在三种不同厂商的3 T扫描仪上,用NIST/ISMRM体模和10名健康志愿者评估多动态多回波序列的定量性能。体模与志愿者的T1、T2、PD测量线性很强,志愿者扫描仪内和扫描仪间变异系数较低;作者结论为该序列对脑弛豫测量和体积测量总体稳健,但对脑组织典型范围外的弛豫值应用需谨慎。

要点

  • 多动态多回波序列在3 T扫描仪上对脑组织T1、T2、PD定量和体积测量总体稳健。
  • 体模扫描仪间变异系数最高为T1 10.86%、T2 15.27%、PD 9.95%,志愿者相应为3.15%、5.76%、3.21%。
  • T1和T2在高于脑组织典型观测值范围时变异趋于增大,对脑外解剖应用需谨慎。

核心结论

  • 多动态多回波序列衍生的定量值在不同厂商3 T扫描仪上对脑弛豫测量和体积测量总体稳健,但对弛豫值超出脑组织典型范围的解剖应用需谨慎。

研究设计与样本

  • 验证研究:使用NIST/ISMRM体模(含标准化T1、T2和质子密度球体)和10名健康志愿者,在三种不同厂商的3 T扫描仪上,于不同日期扫描10次、同一时段扫描2次,采用多动态多回波序列。

诊断标准与关键数值

  • 体模T1、T2、PD测量线性R分别为0.973-1.000、0.979-1.000、0.982-0.999。
  • 志愿者T1、T2、PD测量线性R分别为0.973-1.000、0.979-1.000、0.982-0.999。
  • 体模扫描仪内变异系数最高:T1 2.07%、T2 7.60%、PD 12.86%。
  • 志愿者扫描仪内变异系数最高:T1 1.33%、T2 0.89%、PD 0.77%。
  • 体模扫描仪间变异系数最高:T1 10.86%、T2 15.27%、PD 9.95%;志愿者扫描仪间变异系数最高:T1 3.15%、T2 5.76%、PD 3.21%。

局限性

  • T1和T2在高于脑组织典型观测值范围时变异趋于增大。
  • 作者结论指出,对弛豫值超出脑组织典型范围的解剖应用多动态多回波序列需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to evaluate the linearity, bias, intrascanner repeatability, and interscanner reproducibility of quantitative values derived from a multidynamic multiecho (MDME) sequence for rapid simultaneous relaxometry.

MATERIALS AND METHODS: The NIST/ISMRM (National Institute of Standards and Technology/International Society for Magnetic Resonance in Medicine) phantom, containing spheres with standardized T1 and T2 relaxation times and proton density (PD), and 10 healthy volunteers, were scanned 10 times on different days and 2 times during the same session, using the MDME sequence, on three 3 T scanners from different vendors. For healthy volunteers, brain volumetry and myelin estimation were performed based on the measured T1, T2, and PD. The measured phantom values were compared with reference values; volunteer values were compared with their averages across 3 scanners.

RESULTS: The linearity of both phantom and volunteer measurements in T1, T2, and PD values was very strong (R = 0.973-1.000, 0.979-1.000, and 0.982-0.999, respectively) The highest intrascanner coefficients of variation (CVs) for T1, T2, and PD were 2.07%, 7.60%, and 12.86% for phantom data, and 1.33%, 0.89%, and 0.77% for volunteer data, respectively. The highest interscanner CVs of T1, T2, and PD were 10.86%, 15.27%, and 9.95% for phantom data, and 3.15%, 5.76%, and 3.21% for volunteer data, respectively. Variation of T1 and T2 tended to be larger at higher values outside the range of those typically observed in brain tissue. The highest intrascanner and interscanner CVs for brain tissue volumetry were 2.50% and 5.74%, respectively, for cerebrospinal fluid.

CONCLUSIONS: Quantitative values derived from the MDME sequence are overall robust for brain relaxometry and volumetry on 3 T scanners from different vendors. Caution is warranted when applying MDME sequence on anatomies with relaxometry values outside the range of those typically observed in brain tissue.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Hagiwara A, Hori M, Cohen-Adad J, et al. Linearity, Bias, Intrascanner Repeatability, and Interscanner Reproducibility of Quantitative Multidynamic Multiecho Sequence for Rapid Simultaneous Relaxometry at 3 T: A Validation Study With a Standardized Phantom and Healthy Controls. Invest Radiol. 2019. PMID: 30300164. doi:10.1097/RLI.0000000000000510