彩色多普勒成像与超声造影检测颈动脉粥样硬化病变中溃疡的比较研究
本研究以多排CT血管造影为参考方法,比较彩色多普勒成像(CDI)与超声造影(CEUS)诊断颈动脉斑块溃疡的准确性。纳入54例患者66条颈动脉,CEUS与MDCTA的一致性(kappa 0.921)优于CDI(kappa 0.473),CEUS敏感度94.1%、特异度97.95%,CDI敏感度41.2%、特异度97.95%。
要点
- CEUS诊断颈动脉斑块溃疡的敏感度为94.1%、特异度97.95%,与MDCTA一致性极好(kappa 0.921)。
- CDI诊断溃疡敏感度仅41.2%、特异度97.95%,与MDCTA一致性中等(kappa 0.473)。
- CDI上65.5%的溃疡可见血流反转,CEUS上17.64%可见微气泡漩涡及溃疡旁新生血管。
核心结论
- CEUS在诊断颈动脉溃疡斑块方面,与MDCTA的一致性和诊断准确性均优于CDI。
研究设计与样本
- 以多排CT血管造影(MDCTA)为参考方法,对54例患者66条颈动脉行CDI、CEUS及MDCTA检查,采用kappa统计评价一致性。
关键影像表现/征象
- CDI在65.5%的溃疡中检测到血流反转。
- CEUS在17.64%的溃疡中检测到微气泡漩涡及溃疡旁新生血管。
- MDCTA将28.8%的斑块判为光滑、45.5%为不规则、24.3%为溃疡。
诊断标准与关键数值
- CDI诊断溃疡的敏感度41.2%、特异度97.95%、阳性预测值87.5%、阴性预测值82.8%。
- CEUS诊断溃疡的敏感度94.1%、特异度97.95%、阳性预测值94.1%、阴性预测值97.95%。
- CDI与CEUS诊断溃疡的一致性kappa 0.473,CDI与MDCTA的kappa 0.473,CEUS与MDCTA的kappa 0.921。
临床应用与阅片要点
- 评估颈动脉斑块溃疡时,CEUS与MDCTA一致性极好,可作为超声评估溃疡的补充手段。
- CDI诊断溃疡敏感度有限,阴性结果不能排除溃疡,需结合CEUS或MDCTA进一步评估。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate the diagnostic accuracy of color Doppler imaging (CDI) and contrast-enhanced ultrasound (CEUS) for diagnosing carotid ulceration, having multi-detector computed tomography angiography (MDCTA) as the reference method.
METHODS: Patients with carotid disease referred for ultrasound (US), either due to the occurrence of neurovascular symptoms or for screening purposes, were included in this study if at least one plaque causing moderate (50-69%) or severe (70-99%) internal carotid artery stenosis was detected. Carotid US with CDI technique, CEUS, and MDCTA were performed in all patients, investigating the presence of ulceration. The agreement between modalities was evaluated using kappa statistics.
RESULTS: The study population included 54 patients (median age 62 years, inter-quartile range 16.2) and 66 carotid arteries. The mean degree of stenosis was 68.5% (SD 12.2%) while 47.1% of plaques were symptomatic. MDCTA characterized 28.8% of plaques as smooth, 45.5% irregular, and 24.3% ulcerated. Flow reversal was detected with CDI in 65.5% of ulcerations, while swirling of the microbubbles and neovessels adjacent to the ulcer were detected with CEUS in 17.64%. The agreement for ulceration diagnosis was moderate between CDI and CEUS (kappa 0.473) and between CDI and MDCTA (kappa 0.473) and very good between CEUS and MDCTA (kappa 0.921). The sensitivity, specificity, and positive and negative predictive values of CDI for the diagnosis of ulceration were 41.2%, 97.95%, 87.5%, 82.8% respectively, while CEUS respective measures were 94.1%, 97.95%, 94.1%, and 97.95%.
CONCLUSION: CEUS outperformed CDI in terms of agreement with MDCTA and diagnostic accuracy for the diagnosis of ulcerated carotid plaque.
KEY POINTS: • Superficial ulceration is a significant feature of carotid plaque vulnerability. • Color Doppler imaging has the potential to demonstrate carotid plaque ulceration but is characterized by limited sensitivity and moderate agreement with the reference method of multi-detector computed tomography angiography. • Contrast-enhanced ultrasound outperforms color Doppler imaging in terms of sensitivity for the detection of carotid plaque ulceration and in agreement with the reference method of multi-detector computed tomography angiography.