双层探测器CT胸部、腹部及盆腔扫描采用三分之一碘剂量:图像质量、辐射剂量与最佳单能级设置

Dual-layer detector CT of chest, abdomen, and pelvis with a one-third iodine dose: image quality, radiation dose, and optimal monoenergetic settings.
PubMed 文献 · Clin Radiol · 2018-09-17

本研究纳入40例肾功能不全患者行双层探测器CT胸腹盆扫描,碘剂量为常规的三分之一(200 mg碘/kg),回顾性重建40–70 keV虚拟单能图像,并与40例匹配的常规120 kVp协议患者比较。结果显示DLCT组SSDE约低20%,40–45 keV图像质量与常规120 kVp相当,诊断信心最高。

要点

  • 三分之一碘剂量DLCT胸腹盆CT在40–45 keV单能级下图像质量与常规120 kVp相当。
  • DLCT组SSDE较常规120 kVp组降低约20%(15.4±1.9 vs 19.4±2.3 mGy)。
  • 40–50 keV条纹伪影加重,但40和45 keV诊断信心评分最高且与120 kVp等效。

核心结论

  • 对于碘剂量为常规三分之一的胸腹盆CT,DLCT虚拟单能图像在40–45 keV时可实现辐射剂量降低20%,同时保持与常规120 kVp协议相当的图像质量。

研究设计与样本

  • 回顾性对照研究,DLCT组40例肾功能不全患者(eGFR<45 ml/min/1.73 m2)行减碘剂量胸腹盆CT(120 kVp,200 mg碘/kg),回顾性重建40–70 keV(间隔5 keV)虚拟单能图像;对照组40例匹配患者行常规胸腹盆CT(120 kVp,600 mg碘/kg,迭代重建)。

关键影像表现/征象

  • DLCT虚拟单能图像在整个能量范围内图像噪声几乎恒定(差异≤13%),腹部噪声等于或低于120 kVp。
  • CT衰减值和对比噪声比随能量降低而逐渐升高,在约45–50 keV时达到与120 kVp相当的水平。
  • 40–50 keV时条纹伪影加重(p<0.01),但40和45 keV的诊断信心评分最高,且与120 kVp等效(p=1.0)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To compare the image quality and radiation dose of reduced iodine dose dual-layer detector (DL) computed tomography (CT) with those of a conventional 120 kVp protocol for chest-abdomen-pelvis CT (CAP-CT).

MATERIALS AND METHODS: Forty patients with renal dysfunction (estimated glomerular filtrating ratio <45 ml/min/1.73 m2) underwent reduced iodine dose CAP-CT (120 kVp, 200 mg iodine/kg) on DLCT. Virtual monochromatic images (VMI) at 40-70 keV (5 keV interval) were reconstructed retrospectively. Forty matched patients who underwent conventional CAP-CT (120 kVp, 600 mg iodine/kg, iterative reconstruction) were included as controls. The size-specific dose estimate (SSDE), image noise, CT attenuation, and contrast-to-noise ratio (CNR) were compared between the protocols. Two radiologists rated image contrast, image noise, streak artefact, and diagnostic confidence on a five-point scale.

RESULTS: The SSDE of the DLCT group was approximately 20% lower than that of the 120 kVp group (15.4±1.9 versus 19.4±2.3 mGy, p<0.01). DLCT-VMI provided almost constant image noise throughout the range of energies (differences of ≤13%), with the noise being equivalent or lower than 120 kVp in the abdomen. CT attenuation and CNR gradually increased as the energy decreased, with values comparable to 120 kVp being attained at around 45-50 keV. Although streak artefact was accentuated at 40-50 keV (p<0.01), the highest scores for diagnostic confidence were assigned at 40 and 45 keV, both of which were equivalent to 120 kVp (p=1.0).

CONCLUSION: For CAP-CT with a one-third iodine dose, DLCT-VMI at 40-45 keV allows for a 20% reduction in radiation dose, while preserving image quality comparable to that of conventional 120 kVp protocol.