应用迭代模型重建(IMR)算法的下肢超低辐射剂量CT血管成像

Ultra-low radiation dose CT angiography of the lower extremity using the iterative model reconstruction (IMR) algorithm.
PubMed 文献 · Clin Radiol · 2018-09-06

研究纳入90例患者行256排下肢CTA,分为常规剂量组(120 kVp、自动管电流调制质量指数12、滤波反投影)、低剂量组(80 kVp、质量指数1、IMR)和超低剂量组(80 kVp、20 mAs、IMR)。超低剂量组辐射剂量较常规剂量组降低91.4%、较低剂量组降低67.3%,且超低剂量组与低剂量组客观及主观图像质量均优于常规剂量组,两组间无差异。结论认为使用IMR可在不损害图像质量的前提下将下肢CTA辐射剂量降低达91.4%。

要点

  • 超低剂量组(80 kVp、20 mAs、IMR)辐射剂量0.15±0.02 mSv,较常规剂量组1.86±0.51 mSv降低91.4%。
  • 超低剂量组与低剂量组CT值、SNR、CNR更高、图像噪声更低,主观图像质量更优(均p<0.05)。
  • 超低剂量组与低剂量组之间客观及主观图像质量均无差异(均p>0.05)。

核心结论

  • 下肢CTA使用IMR算法可将辐射剂量降低达91.4%而不损害图像质量。

研究设计与样本

  • 研究论文,90例患者行256排CT下肢CTA,分为常规剂量组、低剂量组和超低剂量组。
  • 常规剂量组:120 kVp、自动管电流调制质量指数12、滤波反投影重建。
  • 低剂量组:80 kVp、自动管电流调制质量指数1、IMR重建。
  • 超低剂量组:80 kVp、20 mAs、IMR重建。

关键影像表现/征象

  • 超低剂量组与低剂量组的CT值、SNR、CNR高于常规剂量组,图像噪声低于常规剂量组(均p<0.05)。
  • 超低剂量组与低剂量组小腿段主观图像质量优于常规剂量组(均p<0.05)。
  • 超低剂量组与低剂量组之间客观及主观图像质量均无差异(均p>0.05)。

诊断标准与关键数值

  • 超低剂量组有效辐射剂量0.15±0.02 mSv,常规剂量组1.86±0.51 mSv,低剂量组0.49±0.08 mSv。
  • 超低剂量组较常规剂量组辐射剂量降低91.4%,较低剂量组降低67.3%(均p<0.05)。

临床应用与阅片要点

  • 下肢CTA采用80 kVp、20 mAs联合IMR重建可在维持图像质量的同时大幅降低辐射剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To investigate the image quality of lower-extremity computed tomography (CT) angiography (CTA) with ultra-low radiation dose using the iterative model reconstruction (IMR) algorithm.

MATERIALS AND METHODS: Lower-extremity CTA was acquired using a 256-multidetector CT system from 90 patients assigned into three groups: (1) the routine dose (RD) group: 120 kVp, automatic tube current modulation (ACTM) with an image quality index of 12, and filtered back projection (FBP); (2) the low-dose (LD) group: 80 kVp, ACTM with an image quality index of 1, and IMR; and (3) the ultra-low dose (ULD) group: 80 kVp, 20 mAs, and IMR. CT attenuation, image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the lower-extremity arteries were calculated. Subjective image quality of lower-extremity segments was assessed. Effective radiation dose was recorded.

RESULTS: The radiation dose was reduced by 91.4% and 67.3% in the ULD group (0.15±0.02 mSv) compared to the RD group (1.86±0.51 mSv) and the LD group (0.49±0.08 mSv; both p<0.05). Higher CT attenuation, SNR, CNR, and lower image noise were obtained in the ULD group and the LD group compared to the RD group (all p<0.05). Better subjective image quality in lower leg segments was obtained in the ULD group and the LD group compared to the RD group (all p<0.05). No difference was found between the ULD and LD groups in both objective and subjective image quality (all p>0.05).

CONCLUSION: By using IMR during lower-extremity CTA, the radiation dose is reduced by up to 91.4% without compromising image quality.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Qian WL, Zhou DJ, Jiang Y, et al. Ultra-low radiation dose CT angiography of the lower extremity using the iterative model reconstruction (IMR) algorithm. Clin Radiol. 2018. PMID: 30195723. doi:10.1016/j.crad.2018.08.001

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。