同步被动声学成像与磁共振测温监测兔肿瘤聚焦超声联合相变纳米乳的空化增强消融

Simultaneous Passive Acoustic Mapping and Magnetic Resonance Thermometry for Monitoring of Cavitation-Enhanced Tumor Ablation in Rabbits Using Focused Ultrasound and Phase-Shift Nanoemulsions.
PubMed 文献 · Ultrasound Med Biol · 2018-09-08

研究在兔肿瘤模型中将相变纳米乳(PSNE)与聚焦超声联用,并在临床磁共振扫描仪内同步进行被动声学成像(PAM)与磁共振测温;PAM可检测纳米乳的汽化与空化,磁共振测温显示空化促进加热,组织学显示含PSNE时热损伤灶更大,94次辐照显示声发射振幅与温升平均增加但辐照间变异显著。

要点

  • PAM可检测相变纳米乳的汽化与空化,并对应磁共振测温所见的加速加热。
  • 磁共振测得的最大加热与积分声发射相关,提示空化增强加热。
  • 含PSNE时组织学热损伤灶更大,与测温数据一致。

核心结论

  • 联合PAM与磁共振可用于监测肿瘤消融,并进一步证明PSNE可促进空化增强的病灶形成。

研究设计与样本

  • 在体兔肿瘤模型研究,在临床磁共振扫描仪内同步进行PAM与磁共振测温,对比使用与不使用PSNE促进空化的条件。

关键影像表现/征象

  • PAM可检测纳米乳的汽化与空化。
  • 含PSNE时磁共振测温显示加热加速。
  • 组织学显示含PSNE时热损伤灶更大,与测温数据一致。

局限性

  • 使用固定辐照条件进行94次辐照显示声发射平均振幅与相应温升增加,但辐照间变异显著,进一步说明需要实时监测。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Thermal ablation of solid tumors via focused ultrasound (FUS) is a non-invasive image-guided alternative to conventional surgical resection. However, the usefulness of the technique is limited in vascularized organs because of convection of heat, resulting in long sonication times and unpredictable thermal lesion formation. Acoustic cavitation has been found to enhance heating but requires use of exogenous nuclei and sufficient acoustic monitoring. In this study, we employed phase-shift nanoemulsions (PSNEs) to promote cavitation and incorporated passive acoustic mapping (PAM) alongside conventional magnetic resonance imaging (MRI) thermometry within the bore of a clinical MRI scanner. Simultaneous PAM and MRI thermometry were performed in an in vivo rabbit tumor model, with and without PSNE to promote cavitation. Vaporization and cavitation of the nanoemulsion could be detected using PAM, which led to accelerated heating, monitored with MRI thermometry. The maximum heating assessed from MRI was well correlated with the integrated acoustic emissions, illustrating cavitation-enhanced heating. Examination of tissue revealed thermal lesions that were larger in the presence of PSNE, in agreement with the thermometry data. Using fixed exposure conditions over 94 sonications in multiple animals revealed an increase in the mean amplitude of acoustic emissions and resulting temperature rise, but with significant variability between sonications, further illustrating the need for real-time monitoring. The results indicate the utility of combined PAM and MRI for monitoring of tumor ablation and provide further evidence for the ability of PSNEs to promote cavitation-enhanced lesioning.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Crake C, Papademetriou IT, Zhang Y, et al. Simultaneous Passive Acoustic Mapping and Magnetic Resonance Thermometry for Monitoring of Cavitation-Enhanced Tumor Ablation in Rabbits Using Focused Ultrasound and Phase-Shift Nanoemulsions. Ultrasound Med Biol. 2018. PMID: 30201425. doi:10.1016/j.ultrasmedbio.2018.07.023

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。