在儿童计算机断层扫描中使用正弦波确认迭代重建的管电流降低程度:体模研究
该体模研究以代表5岁儿童的人体模型扫描头部、胸部及腹部,比较滤波反投影与不同强度正弦波确认迭代重建的图像质量。头部和胸部迭代重建强度3、腹部强度4为最优,对比噪声比提高约50%;头部管电流降低因子为1.2,降低管电流后图像对比噪声比与全管电流滤波反投影图像相似,管电流降低20%并带来相似辐射剂量降低。
要点
- 儿童体模头部和胸部迭代重建最优强度为3,腹部为4,对比噪声比提高约50%。
- 头部管电流降低因子为1.2,对应管电流降低20%并带来相似辐射剂量降低。
- 降低管电流的头部图像对比噪声比与全管电流滤波反投影图像相似。
核心结论
- 已确定头部、胸部及腹部图像迭代重建技术的最优强度;确立管电流降低20%并带来相似辐射剂量降低。
研究设计与样本
- 体模研究,使用代表5岁儿童的人体模型,对头部、胸部及腹部按标准CT协议扫描,图像采用滤波反投影和不同强度正弦波确认迭代重建技术重建。
关键影像表现/征象
- 头部和胸部图像迭代重建最优强度为3,腹部图像最优强度为4。
- 迭代重建最优强度下对比噪声比提高约50%。
- 头部图像管电流降低因子计算为1.2。
- 降低管电流采集的头部图像对比噪声比与全管电流滤波反投影图像相似。
诊断标准与关键数值
- 头部和胸部图像迭代重建最优强度为3。
- 腹部图像迭代重建最优强度为4。
- 头部图像管电流降低因子为1.2,对应管电流降低20%并带来相似辐射剂量降低。
临床应用与阅片要点
- 儿童CT头部和胸部扫描可参考迭代重建强度3、腹部强度4进行参数设置。
- 头部扫描管电流可降低20%并带来相似辐射剂量降低,图像对比噪声比与全管电流滤波反投影图像相似。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Iterative image reconstruction techniques can produce diagnostic-quality computed tomography (CT) images with lower radiation dose.
OBJECTIVE: To quantify the reduction in x-ray tube-current setting and optimize pediatric CT scans using different strengths of an iterative reconstruction technique.
MATERIALS AND METHODS: The head, chest and abdomen regions of an anthropomorphic phantom representing a 5-year-old patient were scanned using standard CT protocols. Images were reconstructed using filtered back projection and different strengths of a sinogram affirmed iterative reconstruction technique. Repeated measurements of contrast-to-noise ratios in the lungs, bone and soft-tissue regions of the phantom were carried out. Maximum increase in contrast-to-noise ratio with iterative reconstruction strength was identified and a tube-current reduction factor was calculated. Head scans were repeated with reduced tube current and compared to filtered back projection images.
RESULTS: Iterative reconstruction strength of 3 for head and chest images and 4 for abdomen images were optimum, resulting in contrast-to-noise ratio increase of about 50%. A tube-current reduction factor of 1.2 for head images was calculated. Images of the head acquired using reduced tube-current showed similar contrast-to-noise ratio as images form filtered back projection with full tube current.
CONCLUSION: Optimum strength of iterative reconstruction technique has been identified for head, chest and abdomen images. Reductions in tube current of 20%, resulting in similar radiation dose reduction, have been established.
引用
- Brindhaban A. Extent of tube-current reduction using sinogram affirmed iterative reconstruction in pediatric computed tomography: phantom study. Pediatr Radiol. 2018. PMID: 30259068. doi:10.1007/s00247-018-4260-3