使用平移最小二乘法从4D数字减影血管造影定量血流速度
本研究提出用平移最小二乘法从4D-DSA数据估算血流速度,在流模体中以超声流量探头为参考验证(相对均方根误差11%),并在14条人体颈内动脉中与相位对比MR比较(Pearson r=0.835)。结果表明4D-DSA可用于估算三维容积内血管的血流速度。
要点
- 4D-DSA可提供时间分辨的三维DSA容积,具有高时间与空间分辨率。
- 平移最小二乘法通过沿中心线的浓度-时间曲线时间平移拟合直线来估算血流速度。
- 流模体验证中4D-DSA平均速度与超声流量探头测量的相对均方根误差为11%。
核心结论
- 4D-DSA使得平移最小二乘法能够在三维容积内估算血管血流速度;流模体与人体颈内动脉验证均显示与参考方法有较好一致性。
研究设计与样本
- 方法学研究:在流模体中以超声流量探头为参考标准验证平移最小二乘法,并在14条人体颈内动脉中与3D相位对比MR成像比较;采用线性回归与相关分析。
关键影像表现/征象
- 从4D-DSA数据集确定所选血管段中心线及各体素沿中心线的浓度-时间曲线。
- 计算沿血管段中心线任一点与其他点之间达到最小差异所需的时间平移。
- 将时间平移作为中心线点位置的函数拟合直线以生成速度。
诊断标准与关键数值
- 流模体:4D-DSA计算的平均速度与超声流量探头测量匹配良好,相对均方根误差11%。
- 人体颈内动脉:4D-DSA计算的平均血流速度与相位对比MR测量相关良好,Pearson相关系数r=0.835。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: 4D-DSA provides time-resolved 3D-DSA volumes with high temporal and spatial resolutions. The purpose of this study is to investigate a shifted least squares method to estimate the blood velocity from the 4D DSA images. Quantitative validation was performed using a flow phantom with an ultrasonic flow probe as ground truth. Quantification of blood velocity in human internal carotid arteries was compared with measurements generated from 3D phase-contrast MR imaging.
MATERIALS AND METHODS: The centerlines of selected vascular segments and the time concentration curves of each voxel along the centerlines were determined from the 4D-DSA dataset. The temporal shift required to achieve a minimum difference between any point and other points along the centerline of a segment was calculated. The temporal shift as a function of centerline point position was fit to a straight line to generate the velocity. The proposed shifted least-squares method was first validated using a flow phantom study. Blood velocities were also estimated in the 14 ICAs of human subjects who had both 4D-DSA and phase-contrast MR imaging studies. Linear regression and correlation analysis were performed on both the phantom study and clinical study, respectively.
RESULTS: Mean velocities of the flow phantom calculated from 4D-DSA matched very well with ultrasonic flow probe measurements with 11% relative root mean square error. Mean blood velocities of ICAs calculated from 4D-DSA correlated well with phase-contrast MR imaging measurements with Pearson correlation coefficient r = 0.835.
CONCLUSIONS: The availability of 4D-DSA provides the opportunity to use the shifted least-squares method to estimate velocity in vessels within a 3D volume.