用于超声评估和表征低至高度可疑甲状腺结节的计算机辅助诊断系统
该回顾性研究纳入62例因甲状腺结节行超声引导下细针穿刺活检的成年患者,比较商用计算机辅助诊断(CAD)系统与放射科医生对结节特征的诊断效能和一致性。CAD系统敏感度显著低于放射科医生(21.4% vs. 78.6%;P=0.008),特异度无统计学差异(81.3% vs. 66.7%;P=0.065)。作者认为该CAD系统是创新工具但需改进诊断性能。
要点
- CAD系统对甲状腺结节特征判读的敏感度显著低于有经验放射科医生(21.4% vs. 78.6%;P=0.008)。
- CAD系统与放射科医生在K-TIRADS分类上一致性仅为一般(K=0.29),边缘判读一致性极低(K=0.03)。
- 特异度方面CAD系统与放射科医生无统计学差异(81.3% vs. 66.7%;P=0.065)。
核心结论
- 该CAD系统敏感度低于有经验放射科医生,与放射科医生对甲状腺结节特征判读的一致性为轻微至中等;作者认为其虽为有良好潜力的创新工具,但仍需进一步努力改进诊断性能。
研究设计与样本
- 回顾性研究,纳入连续系列因甲状腺结节行超声引导下细针穿刺活检的成年患者62例(50例女性;年龄60±12岁),所有患者符合现行国际指南的甲状腺结节细针穿刺活检指征。
诊断标准与关键数值
- CAD系统与放射科医生对甲状腺结节特征判读的观察者间一致性:方向K=0.69(中等),成分K=0.36(一般),回声K=0.36(一般),K-TIRADS K=0.29(一般),边缘K=0.03(轻微)。
- 放射科医生敏感度78.6% vs. CAD系统21.4%(P=0.008),放射科医生显著更高。
- 放射科医生特异度66.7% vs. CAD系统81.3%(P=0.065),无统计学差异。
- 62例甲状腺结节中良性占77.4%(48/62),性质未定至恶性、需随访或手术者占22.6%(14/62)。
临床应用与阅片要点
- 作者认为该CAD系统虽为有良好潜力的创新工具,但仍需进一步努力改进其诊断性能。
局限性
- 原文结论指出该CAD系统虽为有良好潜力的创新工具,但仍需进一步努力改进其诊断性能。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM OF THE STUDY: To compare the diagnostic performance of a commercially available computer-aided diagnosis (CAD) system for thyroid ultrasound (US) with that of a non-computer-aided radiologist in the characterization of low-to-high suspicion thyroid nodules.
METHODS: This retrospective study included a consecutive series of adult patients referred for US-guided fine-needle aspiration biopsy (FNAB) of a thyroid nodule. All patients were eligible for thyroid nodule FNAB according to the current international guidelines. An interventional radiologist experienced in thyroid imaging acquired the US images subsequently used for post-processing, performed FNAB and provided the US features of each nodule. A radiology resident and an endocrinology resident in consensus performed post-processing using the CAD system to assess the same nodule characteristics. The diagnostic performance and agreement of US features between the CAD system and the radiologist were compared.
RESULTS: Sixty-two patients (50 F; age 60 ± 12 years) were enrolled: 77.4% (48/62) of thyroid nodules were benign, 22.6% (14/62) were undetermined to malignant and required follow-up or surgery. Interobserver agreement between the CAD system and the radiologist was substantial for orientation (K = 0.69), fair for composition (K = 0.36), echogenicity (K = 0.36), K-TIRADS (K = 0.29), and slight for margins (K = 0.03). The radiologist demonstrated a significantly higher sensitivity than the CAD system (78.6% vs. 21.4%; P = 0.008), while there was no statistical difference in specificity (66.7% vs. 81.3%; P = 0.065).
CONCLUSION: This CAD system is less sensitive than an experienced radiologist and showed slight-to-substantial agreement with the radiologist for the characterization of thyroid nodules. Although it is an innovative tool with good potential, additional efforts are needed to improve its diagnostic performance.
引用
- Gitto S, Grassi G, De Angelis C, et al. A computer-aided diagnosis system for the assessment and characterization of low-to-high suspicion thyroid nodules on ultrasound. Radiol Med. 2018. PMID: 30244368. doi:10.1007/s11547-018-0942-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。