在上腹部CT中以标称剂量四分之一结合基于模型的迭代重建改善图像质量
研究将40例上腹部CT患者随机分为常规剂量组(噪声指数10 HU,40% ASiR)和低剂量组(噪声指数20 HU,MBIR及40% ASiR),比较延迟期图像。低剂量扫描实现75%剂量降低,MBIR在肝、脾、胰、肾和竖脊肌的SD值低于常规剂量ASiR且CNR更高,主观评分与常规剂量ASiR无差异。
要点
- 低剂量组实现75%剂量降低,MBIR图像质量与常规剂量40% ASiR相当。
- MBIR在肝、脾、胰、肾和竖脊肌的SD值低于常规剂量ASiR,CNR更高。
- 两位医师主观评分一致性良好(κ=0.726)。
核心结论
- 在75%剂量降低条件下,MBIR提供的图像质量与常规剂量下40% ASiR相似。
研究设计与样本
- 随机对照研究,40例上腹部患者分为常规剂量组(n=20,噪声指数10 HU,40% ASiR)和低剂量组(n=20,延迟期噪声指数20 HU,MBIR及40% ASiR),比较延迟期图像。
关键影像表现/征象
- 低剂量组实现75%剂量降低。
- MBIR在肝、脾、胰、肾和竖脊肌的SD值低于常规剂量40% ASiR和低剂量40% ASiR(均p<0.05)。
- MBIR在肝、脾、胰、肾和竖脊肌的CNR值高于常规剂量40% ASiR和低剂量40% ASiR(均p<0.05)。
诊断标准与关键数值
- 主观评分(95%CI):常规剂量40% ASiR组4.00±0.79(3.62-4.37),低剂量40% ASiR组3.35±0.58(3.07-3.62),低剂量MBIR组3.90±0.64(3.60-4.19)。
- 常规剂量40% ASiR组与低剂量MBIR组主观评分无差异,两位医师评分一致性良好(κ=0.726)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the ability of a model-based iterative reconstruction (MBIR) for improving image quality in upper abdominal CT with quarter of the normal dose, in comparison with adaptive statistical iterative reconstruction (ASiR) at normal dose.
METHODS: 40 upper abdominal patients were randomly divided into two groups: normal-dose group (n = 20) with tube current modulation for noise index (NI) of 10 HU and 40% ASiR reconstruction; low-dose group (n = 20) with NI = 20 HU in the delay phase and MBIR and 40%ASiR. Images in the delay phase were compared. The CT values and standard deviation (SD) values of the liver, spleen, pancreas, kidney, erector spine and fat were measured. Contrast-noise-ratio (CNR = (CTtissue-CT fat)/SDfat) of each measured organ were calculated and compared with one-way ANOVA among the three reconstruction groups. The subjective image scores of the three groups were assessed blindly by two experienced physicians using a 5-point system and the score consistency was compared by the κ test.
RESULTS: Dose reduction of 75 % was achieved for the low-dose scan. The subjective scores (95 % confidence intervals) of the three groups (NI 10-40 % ASiR, NI 20-40% ASiR and NI 20-MBIR) were 4.00 ± 0.79 (3.62-4.37), 3.35 ± 0.58 (3.07-3.62) and 3.90 ± 0.64 (3.60-4.19), respectively with no difference between the NI 10-40% ASiR and NI20-MBIR groups and good consistency between reviewers (κ = 0.726). MBIR had statistically lower SD values and higher contrast-to-noise ratio values in the liver, spleen, pancreas, kidney and erector spine than NI 10-40% ASiR and NI 20-40% ASiR (all p < 0.05).
CONCLUSION: At 75 % dose reduction, MBIR provides similar image quality compared to 40% ASiR at normal-dose.
ADVANCES IN KNOWLEDGE: MBIR provides good image quality at 25 % of the normal dose.
引用
- Zhang X, Chen J, Yu N, et al. Improving image quality with model-based iterative reconstruction at quarter of nominal dose in upper abdominal CT. Br J Radiol. 2018. PMID: 30240280. doi:10.1259/bjr.20180137
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。