声辐射力脉冲弹性成像作为活体肝移植供肝无创诊断工具的评价
研究纳入100例受试者(70例受者、30例供者),以肝活检为参考,评估声辐射力脉冲弹性成像(ARFI)预测活体肝移植后移植物显著纤维化(≥F2)的价值,并与瞬时弹性成像(TE)、APRI和FIB-4比较。供者ARFI均值1.05±0.09 m/s;受者中ARFI鉴别≥F2的AUROC为0.93(95%CI 0.86-0.99,P<0.01),最佳截断值1.34 m/s,敏感度90%、特异度82%。ARFI与TE、APRI、FIB-4均显著相关。
要点
- 供者肝脏ARFI速度均值1.05±0.09 m/s,可作为健康供肝的参考值。
- ARFI鉴别肝移植后显著纤维化(≥F2)的AUROC为0.93,最佳截断值1.34 m/s,敏感度90%、特异度82%。
- ARFI与TE、APRI、FIB-4均显著相关(r分别为0.888、0.62、0.585,P均<0.01)。
核心结论
- ARFI可作为评估活体肝移植后显著纤维化的可靠方法。
研究设计与样本
- 共纳入100例受试者(70例受者、30例供者);所有受者计算APRI和FIB-4,所有受试者行TE和ARFI弹性成像(Siemens Acuson S2000超声系统,德国);所有供者和仅30例受者行肝活检;显著纤维化定义为≥F2。
诊断标准与关键数值
- ARFI鉴别≥F2的AUROC为0.93(95%CI 0.86-0.99,P<0.01),最佳截断中位值1.34 m/s,敏感度90%、特异度82%;供者ARFI速度均值1.05±0.09 m/s;受者肝活检纤维化分期<F2者占86.7%,TE分期<F2者占67.1%。
临床应用与阅片要点
- ARFI可用于活体肝移植后显著纤维化的无创评估,供者ARFI均值1.05±0.09 m/s可作为健康肝脏参考;以1.34 m/s为截断值判别≥F2纤维化时敏感度90%、特异度82%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND AIMS: Role of acoustic radiation force impulse (ARFI) elastography, in transplant setting, is not well established. We aimed to define the normal mean values of the liver stiffness by ARFI Elastography in healthy liver donors and to evaluate ARFI elastography as predictor of graft fibrosis post living donor liver transplant (LDLT) in comparison to other non-invasive methods (transient elastography [TE], APRI and FIB4).
PATIENTS AND METHODS: A total of 100 subjects (70 recipients and 30 donors) were recruited. APRI and FIB4 scores were calculated for all recipients. TE and ARFI elastography (Siemens Acuson S2000 Ultrasound System, Germany) were performed to all subjects. All donors and only 30 recipients had liver biopsy. Significant fibrosis was defined as ≥ F2.
RESULTS: The mean ARFI velocity among the donors was 1.05 ± 0.09 m/s. Regarding the recipients: mean age was 49.5 ± 8.49 years, 85.7% males, fibrosis stages < F2 were the most frequent stages by liver biopsy (86.7%) and TE (67.1%). ARFI median was significantly correlated with TE median, APRI and FIB-4 (r = 0.888, p = 0.000; r = 0.62, p = 0.000, and r = 0.585, p = 0.000, respectively). ARFI performed well in discriminating patients with ≥ F2 (AUROC = 0.93, 95% CI 0.86-0.99, p < 0.01) with best cutoff median value of 1.34 m/s (sensitivity 90%, specificity 82%).
CONCLUSION: ARFI can be used as a reliable method in assessment of significant fibrosis post-LDLT.