沸腾组织碎裂术中气泡动力学研究
本研究建立基于Gilmore方程并考虑热质传递的气泡数值模型,结合离体肝实验被动空化检测,分析高强度聚焦超声冲击波激发的气泡动力学。模拟显示激波不对称性与水蒸气输运是气泡在100°C沸腾温度下发生整流生长的关键参数,整流径向运动起始表现为辐射压力升高和频谱中高阶多重谐波突然出现。热消融与沸腾组织碎裂术的频谱对比显示后者出现更高阶多重谐波和更高水平宽带发射,与实验测量一致,可用于监测空化类型及沸腾气泡形成起始。
要点
- 沸腾组织碎裂术频谱中高阶多重谐波与宽带发射升高,与热消融可区分。
- 气泡整流生长起始表现为辐射压力升高和高阶多重谐波突然出现。
- 激波不对称性和水蒸气输运是100°C沸腾温度下整流生长的关键参数。
核心结论
- 激波不对称性与水蒸气输运是气泡在100°C沸腾温度下发生整流生长的关键参数,高阶多重谐波和宽带发射升高可用于监测沸腾组织碎裂术与热消融的不同空化活动及焦点处沸腾气泡形成的起始。
研究设计与样本
- 数值气泡模型基于Gilmore方程并考虑热质传递(气体与水蒸气),研究组织内单气泡在不同高强度聚焦超声场下随流体温度变化的动力学;同时进行离体肝实验,采用被动空化检测系统获取声发射信号。
关键影像表现/征象
- 气泡整流径向运动起始表现为辐射压力升高和相应频谱图中高阶多重谐波突然出现。
- 沸腾组织碎裂术辐照期间出现高阶多重谐波以及更高水平的宽带发射,热消融暴露的频谱则不同。
- 这些声发射特征与实验测量一致。
诊断标准与关键数值
- 模拟显示气泡在沸腾温度100°C下发生整流生长,关键参数为激波不对称性和水蒸气输运。
- 频谱特征:沸腾组织碎裂术出现高阶多重谐波与更高水平宽带发射,热消融暴露未见同样表现。
临床应用与阅片要点
- 可利用高阶多重谐波和宽带发射升高监测沸腾组织碎裂术与热消融的不同空化活动类型。
- 可利用上述声发射特征监测高强度聚焦超声暴露过程中焦点处沸腾气泡形成的起始。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Boiling histotripsy is a non-invasive, cavitation-based ultrasonic technique which uses a number of millisecond pulses to mechanically fractionate tissue. Though a number of studies have demonstrated the efficacy of boiling histotripsy for fractionation of solid tumours, treatment monitoring by cavitation measurement is not well studied because of the limited understanding of the dynamics of bubbles induced by boiling histotripsy. The main objectives of this work are to (a) extract qualitative and quantitative features of bubbles excited by shockwaves and (b) distinguish between the different types of cavitation activity for either a thermally or a mechanically induced lesion in the liver. A numerical bubble model based on the Gilmore equation accounting for heat and mass transfer (gas and water vapour) was developed to investigate the dynamics of a single bubble in tissue exposed to different High Intensity Focused Ultrasound fields as a function of temperature variation in the fluid. Furthermore, ex vivo liver experiments were performed with a passive cavitation detection system to obtain acoustic emissions. The numerical simulations showed that the asymmetry in a shockwave and water vapour transport are the key parameters which lead the bubble to undergo rectified growth at a boiling temperature of 100°C. The onset of rectified radial bubble motion manifested itself as (a) an increase in the radiated pressure and (b) the sudden appearance of higher order multiple harmonics in the corresponding spectrogram. Examining the frequency spectra produced by the thermal ablation and the boiling histotripsy exposures, it was observed that higher order multiple harmonics as well as higher levels of broadband emissions occurred during the boiling histotripsy insonation. These unique features in the emitted acoustic signals were consistent with the experimental measurements. These features can, therefore, be used to monitor (a) the different types of acoustic cavitation activity for either a thermal ablation or a mechanical fractionation process and (b) the onset of the formation of a boiling bubble at the focus in the course of HIFU exposure.
引用
- Pahk KJ, Gélat P, Kim H, et al. Bubble dynamics in boiling histotripsy. Ultrasound Med Biol. 2018. PMID: 30228043. doi:10.1016/j.ultrasmedbio.2018.07.025
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。