应用超声弹性成像测量心室内压力
本研究在8只离体工作兔心上,用压力导管记录左心室心室内压力,同时用剪切波弹性成像测量心肌硬度,发现两者呈线性相关(平均斜率0.318 kPa/mm Hg,R²=0.89),经单点校准后SWEI估测压力与导管测压线性相关(斜率0.95,R²=0.88),提示SWEI弹性成像有可能用于估测心室内压力。
要点
- 离体兔心实验中,剪切波弹性成像测得的心肌硬度与心室内压力呈线性相关(R²=0.89)。
- 经单点校准后,SWEI估测压力与压力导管测压接近等效(斜率0.95,R²=0.88)。
- 作者认为基于压力的指标有可能从SWEI超声弹性成像推导出来。
核心结论
- SWEI测值与美国心室内压力呈线性相关,基于压力的指标有可能从SWEI超声弹性成像推导出来。
- 本研究还显示,采用单点校准用SWEI估测心室内压力具有可行性。
研究设计与样本
- 研究类型为离体工作兔心实验,共8只离体工作兔心。
- 用压力导管记录左心室心室内压力,同时用剪切波弹性成像(SWEI)记录心肌硬度。
- 利用心室内压力与SWEI测得硬度的峰值对SWEI测量进行校准并转换为心室内压力。
诊断标准与关键数值
- 心室内压力与SWEI测值呈过零点的线性关系,平均斜率为0.318 kPa/mm Hg,R²=0.89。
- 用心室内压力/SWEI曲线上的一个点进行校准后,SWEI估测压力与心室内压力线性相关,斜率为0.95,R²=0.88。
- 平均舒张末期压力:压力导管测值为12.716 mm Hg,SWEI估测值为14.726 mm Hg。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Intraventricular pressure (IVP) is one of the most important measurements for evaluating cardiac function, but this measurement is not currently easily assessable in the clinic. The primary reason for this is the absence of a noninvasive technique for measuring IVP. In this study, we investigate the relationship between IVP and dynamic myocardial stiffness measured by shear wave elasticity imaging (SWEI) and assess the feasibility of measuring IVP using SWEI.
METHODS: In 8 isolated working rabbit hearts, IVP was recorded in the left ventricle using a pressure catheter. Simultaneously, myocardial stiffness was recorded by SWEI. Using the peak values for IVP and SWEI measured stiffness, SWEI measurements were calibrated and converted to IVP.
RESULTS: A linear relationship with zero intercept was observed between IVP and SWEI, with the average slope of 0.318 kPa/mm Hg, R2 = 0.89. Using one point on the IVP/SWEI curve, SWEI measurements were converted to IVP. Estimated pressure using SWEI and IVP were linearly correlated with the slope of 0.95, R2 = 0.88 (mean end diastolic pressure by pressure catheter = 12.716 mm Hg and by SWEI=14.726 mm Hg), indicating the near equivalence of the 2 measurements.
CONCLUSION: We have shown that SWEI measurements are linearly related to IVP; therefore, pressure-based indices could potentially be derived from SWEI ultrasound elastography. The feasibility of using SWEI to estimate IVP with a single point calibration was also shown in this study.
引用
- Vejdani-Jahromi M, Freedman J, Trahey GE, et al. Measuring Intraventricular Pressure Using Ultrasound Elastography. J Ultrasound Med. 2018. PMID: 30218456. doi:10.1002/jum.14795
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。