文献速递心血管MR

双源射频发射对3.0 T心脏MR左心室测量及测量可重复性的影响:与传统单源射频发射参考的比较

Effect of Dual-Source Radiofrequency Transmission on Left Ventricular Measurements and Measurement Reproducibility at 3.0 T Cardiac MR Imaging: Comparison with Conventional Single-Source Transmission Reference.
PubMed 文献 · Acad Radiol · 2018-08-30

该前瞻性研究在19名受试者中于3.0 T MR上采用双源射频发射行平衡稳态自由进动电影成像,并与传统单源射频发射参考比较左心室测量值及可重复性。双源射频发射在射血分数上产生小幅但显著的正偏倚,质量显著更小,左心室舒张末期下壁厚度显著低估,但测量可重复性优于单源射频发射。

要点

  • 双源射频发射在3.0 T心脏MR上测得的射血分数较单源射频发射高1.5%±2.6(P=0.021)。
  • 双源射频发射测得的左心室质量较单源射频发射小4.0 g±6.5(P=0.001)。
  • 双源射频发射的观察者内和观察者间变异均低于单源射频发射,变异值<14.0%。

核心结论

  • 双源射频发射在3.0 T MR上改善b-SSFP电影成像图像质量,可能提供比传统单源方法更可重复的左心室测量,并提供合理的左心室测量估计。

研究设计与样本

  • 前瞻性研究,纳入19名受试者,在配备双源射频发射的3.0 T MR上行心脏b-SSFP电影成像,并与传统单源射频发射参考比较;采用配对t检验、线性回归和Bland-Altman分析评估两种技术间差异和一致性,重复测量评估观察者内和观察者间变异。

诊断标准与关键数值

  • 双源射频发射较单源射频发射高估舒张末期容积(平均差3.9 mL±9.7,P>0.05)、收缩末期容积(平均差1.1 mL±2.6,P>0.05)和每搏输出量(平均差2.8 mL±9.1,P>0.05)。
  • 双源射频发射较单源射频发射在射血分数上产生小幅但显著的正偏倚(1.5%±2.6,P=0.021)。
  • 双源射频发射测得的左心室质量显著小于单源射频发射(-4.0 g±6.5,P=0.001)。
  • 双源射频发射较单源射频发射低估室间隔厚度(-0.29 mm±0.6,P=0.067)并显著低估左心室舒张末期下壁厚度(-0.55 mm±0.4,P<0.0001)。
  • 两种技术测量值高度相关(r²=0.81至0.96,P<0.0001);双源射频发射的观察者内和观察者间变异值<14.0%,低于单源射频发射。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To prospectively assess effect of dual-source radiofrequency (RF) transmission on left ventricular (LV) measurements and measurements reproducibility at 3.0 T MR using balanced steady-state free precession (b-SSFP) cine imaging, compared to the conventional single-source RF transmission reference approach.

MATERIALS AND METHODS: Cardiac b-SSFP cine imaging was performed in 19 subjects at 3.0 T MR equipped with dual-source RF transmission. All images were analyzed to obtain LV end-diastolic volume, end-systolic volume, stroke volume, ejection fraction, mass, LV end-diastolic inferior wall thickness, and interventricular septal thickness. The difference of all LV measurements between the two imaging techniques was tested with the paired t test and the intertechnique agreement was tested through linear regression and Bland-Altman analyses. Additionally, repeated LV measurements were performed to determine intra and interobserver variability with the Bland-Altman method, the 95% limits of agreement, the coefficient of variation (CV) and the intraclass correlation coefficient.

RESULTS: Compared to conventional single-source, dual-source slightly overestimated end-diastolic volume, end-systolic volume, and stroke volume (mean differences, 3.9 mL ± 9.7, 1.1 mL ± 2.6, and 2.8 mL ± 9.1, respectively; p > 0.05), resulting in a small but significant positive bias in ejection fraction (1.5% ± 2.6; p = 0.021). Mass was significantly smaller with dual-source than with single-source (-4.0 g ± 6.5, p = 0.001). Dual-source slightly underestimated interventricular septal thickness (-0.29 mm ± 0.6, p = 0.067) and significantly underestimated LV end-diastolic inferior wall thickness (-0.55 mm ± 0.4, p < 0.0001). The two techniques in measurements correlated highly (r2 = 0.81 to 0.96, p < 0.0001). Intra and interobserver variability in dual-source measurements was much lower than that in single-source, and variability values were <14.0%.

CONCLUSION: Improved image quality of b-SSFP cine imaging at 3.0 T MR with dual-source RF transmission may provide more reproducible LV measurements compared to conventional single-source approach. Dual-source RF transmission also provides a reasonable estimate of the LV measurements.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Li C, Liu B, Meng H, et al. Effect of Dual-Source Radiofrequency Transmission on Left Ventricular Measurements and Measurement Reproducibility at 3.0 T Cardiac MR Imaging: Comparison with Conventional Single-Source Transmission Reference. Acad Radiol. 2018. PMID: 30172713. doi:10.1016/j.acra.2018.06.007

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。