三维径向MPnRAGE在未镇静儿童结构MRI中的稳健运动校正策略
本研究开发并评估了一种针对三维径向采集MPnRAGE数据的回顾性运动校正方法。44例未镇静儿童(32例自闭症谱系障碍,12例年龄匹配对照)接受扫描,两名神经放射学家用Likert量表(0-4分)评分,结果显示运动校正使平均Likert评分从3.0提高至3.8,标准差从1.1降至0.4,图像质量显著改善(Mann-Whitney U检验,P < .001)。
要点
- 回顾性运动校正使MPnRAGE T1加权图像平均Likert评分从3.0升至3.8,标准差从1.1降至0.4。
- 校正后所有T1加权图像均被评为良好或优秀,部分病例从不可用(0分)提升至良好(3分)。
- Tenengrad评分与两名阅片者绝对一致性的组内相关系数分别为0.92和0.88(P均< .0005)。
核心结论
- 回顾性运动校正对MPnRAGE数据高度有效,能持续改善未镇静儿童T1加权MRI的图像质量,同时可重建多种组织对比用于结构分析。
研究设计与样本
- 研究为回顾性运动校正方法的开发与评估,采用三维径向数据采集的MPnRAGE技术获取高分辨率容积磁化准备快速梯度回波(MPRAGE)图像及多种组织对比;样本为44例未镇静儿童,其中32例自闭症谱系障碍(平均年龄±标准差,13岁±3)和12例年龄匹配对照(平均年龄,12岁±3)。
诊断标准与关键数值
- 两名经验丰富的神经放射学家采用Likert量表(0-4分,从不可用到优秀)对有无回顾性运动校正的图像进行评分;Tenengrad指标作为无参考的图像锐度测量用于统计分析。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To develop and evaluate a retrospective method to minimize motion artifacts in structural MRI. Materials and Methods The motion-correction strategy was developed for three-dimensional radial data collection and demonstrated with MPnRAGE, a technique that acquires high-resolution volumetric magnetization-prepared rapid gradient-echo, or MPRAGE, images with multiple tissue contrasts. Forty-four pediatric participants (32 with autism spectrum disorder [mean age ± standard deviation, 13 years ± 3] and 12 age-matched control participants [mean age, 12 years ± 3]) were imaged without sedation. Images with and images without retrospective motion correction were scored by using a Likert scale (0-4 for unusable to excellent) by two experienced neuroradiologists. The Tenengrad metric (a reference-free measure of image sharpness) and statistical analyses were performed to determine the effects of performing retrospective motion correction. Results MPnRAGE T1-weighted images with retrospective motion correction were all judged to have good or excellent quality. In some cases, retrospective motion correction improved the image quality from unusable (Likert score of 0) to good (Likert score of 3). Overall, motion correction improved mean Likert scores from 3.0 to 3.8 and reduced standard deviations from 1.1 to 0.4. Image quality was significantly improved with motion correction (Mann-Whitney U test; P < .001). Intraclass correlation coefficients for absolute agreement of Tenengrad scores with reviewers 1 and 2 were 0.92 and 0.88 (P < .0005 for both), respectively. In no cases did the retrospective motion correction induce severe image degradation. Conclusion Retrospective motion correction of MPnRAGE data were shown to be highly effective for consistently improving image quality of T1-weighted MRI in unsedated pediatric participants, while also enabling multiple tissue contrasts to be reconstructed for structural analysis. © RSNA, 2018 Online supplemental material is available for this article.
引用
- Kecskemeti S, Samsonov A, Velikina J, et al. Robust Motion Correction Strategy for Structural MRI in Unsedated Children Demonstrated with Three-dimensional Radial MPnRAGE. Radiology. 2018. PMID: 30063192. doi:10.1148/radiol.2018180180
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。