保持图像质量的骨盆肌肉骨骼CT辐射剂量降低

Radiation dose reduction for musculoskeletal computed tomography of the pelvis with preserved image quality.
PubMed 文献 · Skeletal Radiol · 2018-08-28

该回顾性研究将五台CT扫描仪标准化至最低剂量方案,比较优化前后各约60例骨盆CT的剂量与图像质量。平均CTDIvol和DLP降低约45%,股骨皮质信噪比显著升高,主观图像质量虽有一位观察者评分下降但平均仍达3.3分(5分制),结论为剂量大幅降低后主观与客观图像质量仍保持足够水平。

要点

  • 骨盆CT优化后平均CTDIvol从39±14降至18±12 mGy,DLP从1227±469降至546±384 mGy-cm,均降低约45%。
  • 股骨皮质信噪比在两位观察者中均显著升高(p<0.02),主观图像质量平均3.3分(5分制)维持足够。
  • 低BMI者剂量降低幅度更大,高BMI者DLP更高,与优化前后无关。

核心结论

  • 骨盆CT检查在辐射剂量大幅降低后,主观与客观图像质量仍保持足够。

研究设计与样本

  • 经机构审查委员会批准的回顾性记录审查,豁免知情同意;将五台不同制造商/型号的CT扫描仪通过经验丰富的肌肉骨骼放射科医生对临床图像的主观评估,标准化至匹配 campus 最低剂量方案;分析优化前连续60例和优化后连续59例骨盆骨CT,由两名住院医师标准化测量股骨皮质与骨髓、臀中肌、皮下及内脏脂肪的信噪比,两名盲法主治放射科医生对图像质量进行差至优的评级。

关键影像表现/征象

  • 股骨皮质信噪比在两位观察者中均显著升高(p<0.02)。
  • 主观图像质量在一位观察者中显著下降(p<0.01),但两位观察者的平均质量均保持足够(3.3分,5分制)。

诊断标准与关键数值

  • 信噪比测量观察者间一致性为0.64-0.92。
  • 优化前平均CTDIvol 39±14 mGy vs 优化后18±12 mGy(p<0.0001),降低约45%。
  • 优化前平均DLP 1227±469 mGy-cm vs 优化后546±384 mGy-cm(p<0.0001),降低约45%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To analyze the impact of pelvic computed tomography (CT) technique optimization on estimated dose and subjective and objective image quality.

MATERIALS AND METHODS: An institutional review board (IRB)-approved retrospective records review was performed with waived informed consent. Five CT scanners (various manufacturers/models) were standardized to match the lowest dose profile on campus via subjective assessment of clinical images by experienced musculoskeletal radiologists. The lowest dose profile had previously been established through image assessment by experienced musculoskeletal radiologists after a department-wide radiation dose reduction initiative. A consecutive series of 60 pre- and 59 post-optimization bony pelvis CTs were analyzed by two residents, who obtained signal-to-noise ratio for femoral cortex and marrow, gluteus medius muscle, and subcutaneous and visceral fat in a standardized fashion. Two blinded attending radiologists ranked image quality from poor to excellent.

RESULTS: Pre- and post-optimization subjects exhibited no difference in gender, age, or BMI (p > 0.2). Mean CT dose index (CTDIvol) and dose-length product (DLP) decreased by approximately 45%, from 39± 14 to 18± 12 mGy (p < 0.0001) and 1,227± 469 to 546± 384 mGy-cm (p < 0.0001). Lower body mass index (BMI) was associated with a larger dose reduction and higher BMI with higher DLP regardless of pre- or post-optimization examination. Inter-observer agreement was 0.64-0.92 for SNR measurements. Cortex SNR increased significantly for both observers (p < 0.02). Although qualitative image quality significantly decreased for one observer (p < 0.01), adequate mean quality (3.3 out of 5) was maintained for both observers.

CONCLUSION: Subjective and objective image quality for pelvic CT examination remains adequate, despite a substantially reduced radiation dose.