文献速递骨肌USCT

离体人骨关节炎软骨下骨的超声参数:与微型CT参数的比较

Ultrasound Parameters for Human Osteoarthritic Subchondral Bone ex Vivo: Comparison with Micro-Computed Tomography Parameters.
PubMed 文献 · Ultrasound Med Biol · 2018-07-29

研究对16个离体人骨关节炎股骨外侧髁用15 MHz脉冲回波超声三维扫描,分析软骨-软骨下骨界面区(层1)和内部软骨下骨区(层2)的纹理参数,并与微型CT测量的软骨下孔隙率、软骨下骨板厚度和骨小梁区骨体积分数对照。ENTL1x和ENTL1z与孔隙率相关(r均为0.58),CORL2x与骨板厚度相关(r=-0.73),CORL2z与骨体积分数相关(r=-0.66),其余纹理参数无显著相关。

要点

  • 层1平行方向熵ENTL1x与软骨下孔隙率相关(r=0.58)。
  • 层1垂直方向熵ENTL1z与软骨下孔隙率相关(r=0.58)。
  • 层2平行方向相关CORL2x与软骨下骨板厚度相关(r=-0.73)。

核心结论

  • ENTL1x、ENTL1z、CORL2x和CORL2z是反映骨关节炎相关软骨下骨特征的有效指标,其余纹理参数无显著相关。

研究设计与样本

  • 离体研究,纳入16个人骨关节炎股骨外侧髁,使用15 MHz脉冲回波超声三维扫描系统评估,以微型CT测量的软骨下孔隙率、软骨下骨板厚度和骨小梁区骨体积分数为参考标准。

诊断标准与关键数值

  • ENTL1x与软骨下孔隙率显著相关,r=0.58。
  • ENTL1z与软骨下孔隙率显著相关,r=0.58。
  • CORL2x与软骨下骨板厚度显著相关,r=-0.73。
  • CORL2z与骨小梁区骨体积分数显著相关,r=-0.66。
  • 其他纹理参数与上述参考指标无显著相关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The aim of this study was to identify ultrasound parameters reflecting subchondral porosity (Po), subchondral plate thickness (Tpl) and bone volume fraction at the trabecular bone region (BV/TVTb). Sixteen osteoarthritic human lateral femoral condyles were evaluated ex vivo using a 15-MHz pulsed-echo ultrasound 3-D scanning system. The cartilage-subchondral bone (C-B) surface region (layer 1) and inner subchondral bone region (layer 2) were analyzed; we newly introduced entropy (ENT) and correlation (COR) of ultrasound texture parameters of the parallel (x) or perpendicular (z) direction to the C-B interface for this analysis. Po, Tpl and BV/TVTb were evaluated as reference measurements using micro-computed tomography. ENTL1x (ENT of layer 1, x-direction) and ENTL1z were significantly correlated with Po (both r values = 0.58), CORL2x with Tpl (r = -0.73) and CORL2z with BV/TVTb (r = -0.66). These are efficient indicators of the characteristics of osteoarthritis-related subchondral bone; the other texture parameters were not significant.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Kiyan W, Nakagawa Y, Ito A, et al. Ultrasound Parameters for Human Osteoarthritic Subchondral Bone ex Vivo: Comparison with Micro-Computed Tomography Parameters. Ultrasound Med Biol. 2018. PMID: 30064850. doi:10.1016/j.ultrasmedbio.2018.06.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。