常规120kVp多能CT与70keV单能快速千伏切换双能CT在囊性与实性肾脏肿块中的衰减和强化程度比较
研究比较22例患者39个肾脏肿块(24个Bosniak I/II类囊肿和15个实性肿块)在常规120kVp多能CT与70keV单能双能CT上的衰减值和强化程度,结果显示两种方法在囊肿和实性肿块的绝对衰减及强化程度上均无差异,相关性强至几乎完美。
要点
- 70keV单能CT与常规120kVp CT在肾囊肿和实性肿块的绝对衰减值上无显著差异。
- 两种方法测得的强化程度(衰减变化)在肾囊肿和实性肿块中均无显著差异。
- 两种方法的相关性强至几乎完美(β=0.83-0.98),一致性良好。
核心结论
- 70keV单能CT与常规120kVp CT在肾囊肿和实性肿块的绝对衰减值或强化程度上无差异。
研究设计与样本
- 研究纳入22例患者39个肾脏肿块(24个Bosniak I和II类囊肿及15个实性肿块),均接受常规CT(120kVp平扫和增强)和快速千伏切换双能CT(120kVp平扫和70keV增强),两次扫描平均间隔648±943天,由一名放射科医师在配准图像上测量衰减值。
诊断标准与关键数值
- 囊肿增强后衰减值:120kVp为9.5±5.5 HU(范围-2至20 HU),70keV为10.1±4.6 HU(范围-2至16 HU),p=0.33。
- 实性肿块增强后衰减值:120kVp为110.1±72.9 HU(范围35至267 HU),70keV为119.1±73.7 HU(范围33至265 HU),p=0.04。
- 强化程度(120kVp增强减120kVp平扫):囊肿3.5±3.9 HU(范围-2至13 HU),实性肿块80.7±73.3 HU(范围9至227 HU),p=0.45。
- 强化程度(70keV增强减120kVp平扫):囊肿4.3±4.1 HU(范围-3至12 HU),实性肿块89.8±74.1 HU(范围7至226 HU),p=0.04。
- 两种方法相关性β=0.83-0.98,一致性良好;不同时间点120kVp采集的囊肿和实性肿块衰减值无差异(p=0.20-0.92),相关性β=0.72-0.95。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study was to compare attenuation values (in Hounsfield units) and degree of enhancement (attenuation change) in renal masses using 120-kVp polychromatic (conventional) CT and 70-keV monochromatic dual-energy CT (DECT).
MATERIALS AND METHODS: Twenty-two patients with 39 renal masses (24 Bosniak category I and II cysts and 15 solid masses under active surveillance) underwent conventional CT (120-kVp unenhanced and contrast-enhanced CT) and rapid kilovoltage-switching DECT (120-kVp unenhanced CT and 70-keV contrast-enhanced CT). The mean (± SD) time between scans was 648 ± 943 days. A radiologist measured attenuation on matched image sets coregistered between examinations. Absolute attenuation and attenuation change were compared using independent t tests, Pearson correlation, and Bland-Altman analysis.
RESULTS: There was no difference in attenuation on 120-kVp versus 70-keV contrast-enhanced CT images for cysts (9.5 ± 5.5 HU [range, -2 to 20 HU] vs 10.1 ± 4.6 HU [range, -2 to 16 HU]; p = 0.33) and solid masses (110.1 ± 72.9 HU [range, 35-267 HU] vs 119.1 ± 73.7 HU [range, 33-265 HU]; p = 0.04). There also was no difference in attenuation change for 120-kVp contrast-enhanced CT minus 120-kVp unenhanced CT (cysts, 3.5 ± 3.9 HU [range, -2 to 13 HU]; solid masses, 80.7 ± 73.3 HU [range, 9-227 HU]; p = 0.45) or for 70-keV contrast-enhanced CT minus 120-kVp unenhanced CT (cysts, 4.3 ± 4.1 HU [range, -3 to 12 HU]; solid masses, 89.8 ± 74.1 HU [range, 7-226 HU]; p = 0.04). The correlation was strong to almost perfect (β = 0.83-0.98) with substantial agreement. There was no difference in attenuation of cysts and solid masses comparing 120-kVp acquisitions acquired at different time points (p = 0.20-0.92). The correlation was strong to almost perfect (β = 0.72-0.95) with substantial agreement.
CONCLUSION: There are no differences in absolute attenuation or degree of enhancement comparing 70-keV monochromatic CT to conventional 120-kVp CT in renal cysts and solid masses.