超声剪切波弹性成像:深度、力度及距换能器中心轴距离对肝纤维化评估的影响及不同系统比较
前瞻性纳入20例弥漫性肝病患者,用瞬时弹性成像、ARFI和2-D-SWE测量剪切波速度,评估深度、预载力和离轴距离对纤维化分期的影响。深度是唯一影响ARFI测值的因素,2-D-SWE测值不受采集因素影响;在7cm深度沿中心轴诊断显著纤维化时,ARFI和2-D-SWE的AUC分别为0.92和0.82,与瞬时弹性成像相当。
要点
- 深度是唯一影响ARFI测值的采集因素,2-D-SWE测值不受深度、预载力及离轴距离影响。
- 7cm深度沿中心轴诊断显著纤维化(F≥2)时,ARFI的AUC为0.92,2-D-SWE的AUC为0.82。
- ARFI和2-D-SWE在7cm深度沿中心轴的诊断效能与瞬时弹性成像相当。
核心结论
- 深度是唯一影响ARFI衍生测值的因素,2-D-SWE的剪切波速度值不受所评估采集因素影响;在7cm深度沿中心轴诊断显著纤维化时,ARFI和2-D-SWE均表现出良好诊断效能,与瞬时弹性成像相当。
研究设计与样本
- 前瞻性研究,纳入20例弥漫性肝病患者;使用瞬时弹性成像、声辐射力脉冲(ARFI)和2-D剪切波弹性成像(2-D-SWE)测量剪切波速度;ARFI和2-D-SWE在不同深度(3、5、7cm)、不同预载力(4、7、10N及可变力)和距换能器中心轴0、2、4cm处测量;由一名盲法病理学家采用METAVIR系统(F0-F4)进行纤维化分期。
诊断标准与关键数值
- 以METAVIR F≥2为显著纤维化,在7cm深度沿中心轴测量时,ARFI诊断显著纤维化的AUC为0.92。
- 以METAVIR F≥2为显著纤维化,在7cm深度沿中心轴测量时,2-D-SWE诊断显著纤维化的AUC为0.82。
- ARFI和2-D-SWE在7cm深度沿中心轴诊断显著纤维化的诊断效能与瞬时弹性成像相当。
局限性
- 原文结论指出,进一步研究这一发现可能会引起兴趣。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
We evaluated variation in fibrosis staging caused by depth, pre-load force and measurement off-axis distance on different ultrasound shear wave elastography (SWE) systems prospectively in 20 patients with diffuse liver disease. Shear wave speed (SWS) was measured with transient elastography, acoustic radiation force impulse (ARFI) and 2-D shear wave elastography (SWE). ARFI and 2-D-SWE measurements were obtained at different depths (3, 5 and 7 cm), with different pre-load forces (4, 7 and 10N and variable) and at 0, 2 and 4cm off the central axis of the transducer. A single, blinded pathologist staged fibrosis using the METAVIR system (F0-F4). Area under the receiver operating characteristic curve was charted to differentiate significant fibrosis (F ≥ 2). Depth was the only factor found to influence ARFI-derived values; no acquisition factors were found to affect 2-D-SWE SWS values. ARFI and 2-D-SWE for diagnosis of significant fibrosis at a depth of 7cm along the central axis had good diagnostic performance (areas under the receiver operating characteristic curve: 0.92 and 0.82, respectively), comparable to that of transient elastography. Further investigation of this finding will likely be of interest.
引用
- Dhyani M, Xiang F, Li Q, et al. Ultrasound Shear Wave Elastography: Variations of Liver Fibrosis Assessment as a Function of Depth, Force and Distance from Central Axis of the Transducer with a Comparison of Different Systems. Ultrasound Med Biol. 2018. PMID: 30143339. doi:10.1016/j.ultrasmedbio.2018.07.003