使用自动三维定量算法测量存在节段性室壁运动异常患者的左心室容积与射血分数

Measurement of Left Ventricular Volumes and Ejection Fraction in Patients with Regional Wall Motion Abnormalities Using an Automated 3D Quantification Algorithm.
PubMed 文献 · Ultrasound Med Biol · 2018-08-17

研究纳入72例存在室壁运动异常的患者,以手动三维定量为标准,比较HeartModel三种模式测量左心室舒张末期容积、收缩末期容积和射血分数的准确性与耗时。区域心内膜边界编辑模式与手动三维相关性最佳(r分别为0.88、0.93、0.91),耗时67.3±13.0秒,偏差较小且一致性界限窄。左心室室壁运动异质性与该模式测量射血分数和手动三维的差值相关(P=0.014,风险比=5.19)。

要点

  • 对存在室壁运动异常的患者,HeartModel区域心内膜边界编辑模式测量左心室容积和射血分数与手动三维定量相关性最佳。
  • 区域编辑模式测量左心室舒张末期容积、收缩末期容积和射血分数的偏差分别为11.7 mL、8.45 mL和-1.57%。
  • 左心室室壁运动异质性与区域编辑模式测量射血分数和手动三维的差值相关(P=0.014)。

核心结论

  • 在存在室壁运动异常的患者中,使用区域心内膜边界编辑的HeartModel可准确、高效地评估左心室容积和射血分数。

研究设计与样本

  • 研究纳入72例存在室壁运动异常的患者,使用EPIQ超声诊断仪(X5-1探头,Philips Healthcare)行超声心动图检查,以常规二维Simpson双平面法和手动三维定量(QLAB-3DQA软件,Philips Healthcare)作为标准方法进行对比。

关键影像表现/征象

  • HeartModel三种模式中,区域心内膜边界编辑模式与手动三维定量评估左心室舒张末期容积、收缩末期容积和射血分数的相关性最佳(r分别为0.88、0.93和0.91)。
  • 区域心内膜边界编辑模式测量左心室舒张末期容积、收缩末期容积和射血分数的偏差分别为11.7 mL、8.45 mL和-1.57%,一致性界限窄。
  • 区域心内膜边界编辑模式耗时略长,为67.3±13.0秒。

诊断标准与关键数值

  • 以手动三维定量(QLAB-3DQA软件)作为标准方法,区域心内膜边界编辑模式测量左心室舒张末期容积、收缩末期容积和射血分数的相关系数分别为r=0.88、0.93和0.91。
  • 区域心内膜边界编辑模式测量左心室舒张末期容积、收缩末期容积和射血分数的偏差分别为11.7 mL、8.45 mL和-1.57%。
  • 左心室室壁运动异质性与区域心内膜边界编辑模式和手动三维测量射血分数的差值相关(P=0.014,风险比=5.19)。

临床应用与阅片要点

  • 对存在节段性室壁运动异常的患者,可采用HeartModel区域心内膜边界编辑模式测量左心室舒张末期容积、收缩末期容积和射血分数,其与手动三维定量相关性最佳。
  • HeartModel区域心内膜边界编辑模式测量耗时67.3±13.0秒,略长于其他模式。
  • 左心室室壁运动异质性与区域心内膜边界编辑模式测量射血分数和手动三维的差值相关,提示在室壁运动异质性明显的患者中需关注该模式测量射血分数的偏差。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Accurate and rapid left ventricular (LV) ejection fraction (EF) measurement is crucial for patients with wall motion abnormalities (WMAs). Conventional 2D echocardiographic imaging has limitations. The recently developed software HeartModel (HM, Philips Healthcare, Andover, MA, USA) has shown promise in automated 3D quantification. However, the accuracy and detailed features of HM in measurements of LV volume and EF in patients with regional WMAs have not been carefully investigated. In the present study, echocardiographic imaging (EPIQ, X5-1, Philips Healthcare) was performed in 72 patients with WMAs. The LV end-diastolic volume (EDV), end-systolic volume (ESV) and EF were measured by HM in three modes: without editing and with global and regional endocardial border editing (Auto 3D-NE, Auto 3D-GE and Auto 3D-RE, respectively). The conventional 2D Simpson's biplane method and manual 3D quantification (QLAB-3DQA software, Philips Healthcare), as the standard method, were used for comparison. Among the three HM modalities, Auto 3D-RE exhibited the best correlation with manual 3D in assessing EDV, ESV and EF (r = 0.88, 0.93 and 0.91, respectively), although it took slightly longer (67.3 ± 13.0 s). Auto 3D-RE also exhibited a small degree of bias for the measurements (EDV: 11.7mL, ESV: 8.45mL, EF: -1.57%) and narrow limits of agreement. Heterogeneity of LV wall motion was defined to indicate the dispersion degree of WMAs. It associated with the difference in EF measurement between Auto 3D-RE and manual 3D (p = 0.014, hazard ratio = 5.19). In patients with WMAs, HM with regional contour editing enables accurate and efficient evaluation of LV volume and EF.