CEUS与CT血管造影在腹主动脉内支架随访中的诊断准确性及基于活动的成本分析
该回顾性研究纳入137例腹主动脉瘤腔内修复术后随访患者,以CTA为参考标准评估CEUS诊断内漏的准确性,并做基于活动的成本分析。CEUS诊断内漏的准确率97.4%、敏感度96%、特异度100%、阳性预测值100%、阴性预测值93.1%;作者认为CEUS准确且便宜,在术后第一年阴性对照后可考虑作为CTA的有效替代。
要点
- CEUS诊断内漏的敏感度96%、特异度100%、准确率97.4%,以CTA为参考标准。
- CEUS成本84.7欧元 vs CTA成本157.77欧元,差价73.07欧元。
- 作者认为术后第一年阴性对照后,CEUS可考虑作为CTA的有效替代。
核心结论
- CEUS是腹主动脉瘤腔内修复术后随访中准确且便宜的影像方法,在术后第一年阴性对照后可考虑作为CTA的有效替代,减少不必要的CT检查。
研究设计与样本
- 回顾性研究,纳入137例腹主动脉瘤腔内修复术后随访患者,随访期6年(术后平均随访无动脉瘤囊体积缩小)。
- 以CT血管造影(CTA)为参考标准,计算CEUS的敏感度、特异度、阳性预测值、阴性预测值和准确率。
- 进行基于活动的成本分析,评估在术后第一年阴性对照后以CEUS替代CT的潜在节省。
诊断标准与关键数值
- CEUS检测和表征内漏的准确率97.4%、敏感度96%、特异度100%、阳性预测值100%、阴性预测值93.1%,以CTA为参考标准。
- CEUS成本84.7欧元,CTA成本157.77欧元,差价73.07欧元。
临床应用与阅片要点
- 作者认为在术后第一年阴性对照后,CEUS可考虑作为CTA的有效替代,以减少不必要的CT检查。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate diagnostic accuracy and to perform an activity-based cost analysis of contrast-enhanced ultrasonography (CEUS) compared to computed tomography (CT) during annual surveillance after abdominal aortic aneurysm repair with endovascular procedure (EVAR).
MATERIALS AND METHODS: This retrospective study included 137 patients in post-EVAR follow-up over a 6-year period (average post-operatory follow-up without aneurysm sac volumetric reduction). Sensitivity, specificity, positive predictive values, negative predictive values and accuracy were considered for CEUS using CT angiography (CTA) as reference standard. An activity-based cost analysis was performed to evaluate potential savings due to the introduction of CEUS as an alternative to CT, after the first year of postoperative negative controls.
RESULTS: CEUS reported accuracy, sensitivity, specificity, positive predictive values, negative predictive values of 97.4, 96, 100, 100 and 93.1% in the detection and characterization of endoleaks. CEUS cost was € 84.7, and CTA cost was € 157.77, with a differential cost of € 73.07; using CEUS as an alternative to CT allowed a potential saving of 50.052,95 € during follow-up.
CONCLUSIONS: CEUS is an accurate and cheap imaging method in post-EVAR follow-up patients, and it could be considered as a valid alternative to CTA, after the first year of negative controls, reducing the number of unnecessary CT examinations.
引用
- Faccioli N, Foti G, Casagranda G, et al. CEUS versus CT Angiography in the follow-up of abdominal aortic endoprostheses: diagnostic accuracy and activity-based cost analysis. Radiol Med. 2018. PMID: 30084107. doi:10.1007/s11547-018-0926-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。