用于儿童三维经食管超声心动图的微型矩阵换能器的声学特性
本文介绍一种用于儿童实时三维经食管超声心动图(TEE)的微型PZT-on-CMOS矩阵换能器的设计、制造与声学特性。该换能器采用32×32阵列与分区的收发架构,中心频率5.8 MHz,-6 dB带宽67%,发射效率6 kPa/V(30 mm处),接收动态范围85 dB,最小与最大可检测压力分别为5 Pa和84 kPa,作者认为其特性非常适合微型儿童实时三维TEE探头。
要点
- 中心频率5.8 MHz、-6 dB带宽67%,适合儿童三维TEE的声学需求
- 发射效率6 kPa/V(30 mm处),接收动态范围85 dB
- 最小与最大可检测压力分别为5 Pa和84 kPa
核心结论
- 该PZT-on-CMOS矩阵换能器具有5.8 MHz中心频率、67%的-6 dB带宽、30 mm处6 kPa/V的发射效率及85 dB接收动态范围,作者认为其特性非常适合微型儿童实时三维TEE探头。
研究设计与样本
- 设计、制造与声学特性表征研究;换能器为32×32 PZT-on-CMOS矩阵阵列,集成于专用集成电路(ASIC)之上,采用分区收发架构,中心8×8发射阵元直接引出,其余接收阵元分为96个3×3子阵列,回波由ASIC内微型波束形成电路局部处理,可预偏转至±37°。
诊断标准与关键数值
- 中心频率5.8 MHz,-6 dB带宽67%
- 30 mm处发射效率6 kPa/V
- 接收动态范围85 dB,最小可检测压力5 Pa,最大可检测压力84 kPa
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This paper presents the design, fabrication and characterization of a miniature PZT-on-CMOS matrix transducer for real-time pediatric 3-dimensional (3D) transesophageal echocardiography (TEE). This 3D TEE probe consists of a 32 × 32 array of PZT elements integrated on top of an Application Specific Integrated Circuit (ASIC). We propose a partitioned transmit/receive array architecture wherein the 8 × 8 transmitter elements, located at the centre of the array, are directly wired out and the remaining receive elements are grouped into 96 sub-arrays of 3 × 3 elements. The echoes received by these sub-groups are locally processed by micro-beamformer circuits in the ASIC that allow pre-steering up to ±37°. The PZT-on-CMOS matrix transducer has been characterized acoustically and has a centre frequency of 5.8 MHz, -6 dB bandwidth of 67%, a transmit efficiency of 6 kPa/V at 30 mm, and a receive dynamic range of 85 dB with minimum and maximum detectable pressures of 5 Pa and 84 kPa respectively. The properties are very suitable for a miniature pediatric real-time 3D TEE probe.
引用
- Daeichin V, Bera D, Raghunathan S, et al. Acoustic characterization of a miniature matrix transducer for pediatric 3D transesophageal echocardiography. Ultrasound Med Biol. 2018. PMID: 30072206. doi:10.1016/j.ultrasmedbio.2018.06.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。