纸基3D打印仿真CT体模:两种构建技术的可行性

Paper-based 3D printing of anthropomorphic CT phantoms: Feasibility of two construction techniques.
PubMed 文献 · Eur Radiol · 2018-08-16

研究将两例患者CT图像用碘化钾水溶液(0.6 g/ml)打印在纸上,开发两种方法组装头颈部仿真体模:纸基3D打印机复现外部形状精确但可见胶水伪影,墨粉粘合避免伪影且体模尺寸更灵活。应用研究中两种体模在五种管电流和四种重建设置下,AIDR 3D较FBP获得非劣的SNR和CNR,DLP最高降低83.7%。

要点

  • 纸基3D打印体模可复现患者解剖细节,用于CT剂量与图像质量测试。
  • 3D打印机法外部形状精确但可见胶水伪影,墨粉粘合法无此伪影且尺寸更灵活。
  • AIDR 3D较FBP在体模上获得非劣SNR和CNR,DLP最高降低83.7%。

核心结论

  • 两种将阻射打印纸片组装为个体化患者体模的方法可行,样本应用显示可用于模拟患者成像及系统性CT剂量与图像质量评估。

研究设计与样本

  • 可行性研究,基于两例患者CT图像以碘化钾水溶液(0.6 g/ml)打印于纸上,开发两种组装头颈部体模的方法,并在五种管电流设置下用FBP和AIDR 3D(三种强度)重建,采用双因素方差分析。

关键影像表现/征象

  • 两种方法均获得具有详细患者解剖结构的仿真体模。
  • 3D打印机法精确复现患者外部形状,但可见胶水伪影。
  • 墨粉粘合法避免胶水伪影,且体模尺寸灵活性更高。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To develop and evaluate methods for assembling radiopaque printed paper sheets to realistic patient phantoms for CT dose and image quality testing.

METHODS: CT images of two patients were radiopaque printed with aqueous potassium iodide solution (0.6 g/ml) on paper. Two methods were developed for assembling the paper sheets to head and neck phantoms. (1) Printed sheets were fed to a paper-based 3D printer along with corresponding 3D printable STL files. (2) Paper stacks of 5-mm thickness were glued with toner, cut to the patient shape and assembled to a phantom. In a sample application study, both phantoms were examined with five different tube current settings. Images were reconstructed using filtered-back projection (FBP) and iterative reconstruction (AIDR 3D) with three strength levels. Dose length product (DLP), signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNRs) were analysed. Data were analysed using 2-way analysis of variance (ANOVA).

RESULTS: Both methods achieved anthropomorphic phantoms with detailed patient anatomy. The 3D printer yielded a precise reproduction of the external patient shape, but caused visible glue artefacts. Gluing with toner avoided these artefacts and yielded more flexibility with regard to phantom size. In the sample application study, non-inferior SNR and CNR and up to 83.7% lower DLP were achieved on the phantoms with AIDR 3D compared with FBP.

CONCLUSIONS: Two methods for assembling radiopaque printed paper sheets to phantoms of individual patients are presented. The sample application demonstrates potential for simulation of patient imaging and systematic CT dose and image quality assessment.

KEY POINTS: • Two methods were developed to create realistic CT phantoms of individual patients from radiopaque printed paper sheets. • Analysis of five tube current and four reconstruction settings on two radiopaque 3D printed patient phantoms yielded non-inferior SNR and CNR and up to 83.7% lower dose with iterative reconstruction in comparison with filtered back projection. • Radiopaque 3D printed phantoms can simulate patients and allow systematic analysis of CT dose and image quality parameters.