文献速递神经外周血管CT

单能金属伪影减少技术用于减少介入后脑CT及CT血管成像中的弹簧圈金属伪影

Single-energy metal artifact reduction technique for reducing metallic coil artifacts on post-interventional cerebral CT and CT angiography.
PubMed 文献 · Neuroradiology · 2018-08-24

本研究纳入20例颅内动脉瘤弹簧圈栓塞术后患者,行320排脑CT及CTA,比较应用与未应用SEMAR算法的图像。结果显示SEMAR CT图像伪影强度显著低于标准CT(25.7±10.2 H.U. vs 80.4±67.2 H.U.,p<0.01),且弹簧圈周围解剖结构、血管显示及动脉瘤栓塞状态显示均显著改善。

要点

  • SEMAR可显著降低弹簧圈金属伪影,伪影强度25.7±10.2 H.U. vs 标准CT 80.4±67.2 H.U.(p<0.01)。
  • SEMAR改善弹簧圈周围解剖结构、血管及动脉瘤栓塞状态的显示(p<0.01)。
  • 评估采用双放射科医师独立盲法定性评分及伪影强度定量测量。

核心结论

  • SEMAR算法显著减少颅内动脉瘤弹簧圈栓塞术后脑CT的金属伪影,改善弹簧圈周围解剖结构、血管及动脉瘤栓塞状态的显示。

研究设计与样本

  • 对20例颅内动脉瘤弹簧圈栓塞术后患者使用320排CT行脑CT及CTA,每例重建应用与未应用SEMAR算法的图像(SEMAR CT与标准CT),由两名放射科医师独立盲法定性评估,并以弹簧圈周围衰减值标准差定量评估伪影强度。

关键影像表现/征象

  • SEMAR CT图像伪影强度为25.7±10.2 H.U.,标准CT图像为80.4±67.2 H.U.,SEMAR显著低于标准CT(p<0.01,配对t检验)。
  • SEMAR CT较标准CT显著改善弹簧圈周围解剖结构的显示(p<0.01,符号检验)。
  • SEMAR CT较标准CT显著改善弹簧圈周围动脉的勾勒(p<0.01)。
  • SEMAR CT较标准CT显著改善弹簧圈栓塞动脉瘤状态的显示(p<0.01)。

诊断标准与关键数值

  • 伪影强度以弹簧圈周围衰减值的标准差(H.U.)定量评估。
  • 图像质量由两名放射科医师独立盲法定性评估,包括弹簧圈周围解剖结构显示、弹簧圈周围动脉勾勒及弹簧圈栓塞动脉瘤状态显示。

临床应用与阅片要点

  • 对颅内动脉瘤弹簧圈栓塞术后行脑CT及CTA时,SEMAR算法可减少弹簧圈金属伪影并改善弹簧圈周围结构及血管的显示。
  • 图像评估采用双医师独立盲法,伪影强度以弹簧圈周围衰减值标准差定量测量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the effects of the single-energy metal artifact reduction (SEMAR) algorithm on image quality of cerebral CT and CT angiography (CTA) for patients who underwent intracranial aneurysm coiling.

METHODS: Twenty patients underwent cerebral CT and CTA using a 320-detector row CT after intracranial aneurysm coiling. Images with and without application of the SEMAR algorithm (SEMAR CT and standard CT images, respectively) were reconstructed for each patient. The images were qualitatively assessed by two independent radiologists in a blinded manner for the depiction of anatomical structures around the coil, delineation of the arteries around the coil, and the depiction of the status of coiled aneurysms. Artifact strength was quantitatively assessed by measuring the standard deviation of attenuation values around the coil.

RESULTS: The strength of artifacts measured in SEMAR CT images was significantly lower than that in standard CT images (25.7 ± 10.2 H.U. vs. 80.4 ± 67.2 H.U., p < 0.01, Student's paired t test). SEMAR CT images were significantly improved compared with standard CT images in the depiction of anatomical structures around the coil (p < 0.01, the sign test), delineation of the arteries around the coil (p < 0.01), and the depiction of the status of coiled aneurysms (p < 0.01).

CONCLUSION: The SEMAR algorithm significantly reduces metal artifacts from intracranial aneurysm coiling and improves visualization of anatomical structures and arteries around the coil, and depiction of the status of coiled aneurysms on post-interventional cerebral CT.

来源:Neuroradiology,查看原文。版权归原作者所有。