利用声电成像选择性标测脑深部电刺激导线电流
该概念验证研究将1 MHz超声聚焦于临床脑深部电刺激(DBS)装置附近,在经不同导线注入短电流脉冲时,用记录电极检测高频声电相互作用信号,并通过超声束线性扫描生成电流幅度和极性的时变图像。结果显示可在0.70 mm内精确定位DBS导线,检测阈值为1 MPa下1.75 mA,灵敏度为0.52 ± 0.07 μV/(mA*MPa);经人颅骨重复单极和偶极配置时,尽管骨衰减13.8 dB,信号仍高于背景>10 dB,灵敏度为0.56 ± 0.10 μV/(mA*MPa)。
要点
- 声电成像可非侵入性生成DBS导线电流的幅度与极性时变图像。
- 在盐水中DBS导线定位精度达0.70 mm,检测阈值1.75 mA(1 MPa下)。
- 经人颅骨后信号仍高于背景>10 dB,骨衰减为13.8 dB。
核心结论
- 该概念验证研究表明,利用非侵入性声电成像可能实现对DBS装置导线电流的选择性标测。
研究设计与样本
- 概念验证研究,使用1 MHz超声聚焦于临床DBS装置附近,在经不同DBS导线注入短电流脉冲时,用记录电极检测高频声电相互作用信号,并通过超声束线性扫描生成图像;在盐水中及经人颅骨重复单极和偶极配置。
关键影像表现/征象
- 线性扫描超声束可产生电流幅度和极性随时间变化的图像。
- 盐水中DBS导线定位精度在0.70 mm以内。
- 经人颅骨后,尽管超声衰减13.8 dB,声电成像信号仍高于背景>10 dB。
诊断标准与关键数值
- 盐水中检测阈值为1 MPa下1.75 mA。
- 盐水中灵敏度为0.52 ± 0.07 μV/(mA*MPa)。
- 经人颅骨后灵敏度为0.56 ± 0.10 μV/(mA*MPa)。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
We describe a new application of acoustoelectric imaging for non-invasive mapping of the location, magnitude and polarity of current generated by a clinical deep brain stimulation (DBS) device. Ultrasound at 1MHz was focused near the DBS device as short current pulses were injected across different DBS leads. A recording electrode detected the high-frequency acoustoelectric interaction signal. Linear scans of the US beam produced time-varying images of the magnitude and polarity of the induced current, enabling precise localization of the DBS leads within 0.70mm, a detection threshold of 1.75mA at 1 MPa and a sensitivity of 0.52 ± 0.07 μV/(mA*MPa). Monopole and dipole configurations in saline were repeated through a human skullcap. Despite 13.8-dB ultrasound attenuation through bone, acoustoelectric imaging was still >10dB above background with a sensitivity of 0.56 ± 0.10 μV/(mA*MPa). This proof-of-concept study indicates that selective mapping of lead currents through a DBS device may be possible using non-invasive acoustoelectric imaging.
引用
- Preston C, Kasoff WS, Witte RS, et al. Selective Mapping of Deep Brain Stimulation Lead Currents Using Acoustoelectric Imaging. Ultrasound Med Biol. 2018. PMID: 30119863. doi:10.1016/j.ultrasmedbio.2018.06.021
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。