迭代算法减少儿童骨科假体金属伪影的初步结果
本研究纳入13例11–17岁带金属植入物的儿童,行临床常规CT并用iMAR算法重建,与常规ADMIRE算法比较。近植入物处iMAR较ADMIRE显著改变骨、肌肉、脂肪的CT值(P<0.001),骨噪声显著降低(P<0.003);远植入物处两者无显著差异。
要点
- 近金属植入物处iMAR可显著改变骨、肌肉、脂肪CT值并降低骨噪声。
- 远植入物处iMAR与常规ADMIRE的组织CT值和噪声无显著差异。
- 提示iMAR可在不改变远场基线组织密度和噪声的前提下改善近场图像质量。
核心结论
- 初步结果显示iMAR算法改善低剂量儿童CT近植入物处的CT值并降低骨噪声,且不改变远植入物处的基线组织密度或噪声。
研究设计与样本
- 对13例11–17岁带金属植入物的儿童行临床常规CT,数据分别用iMAR和常规第三代ADMIRE算法重建,在近植入物条纹伪影处和远植入物处测量骨、肌肉、脂肪的CT值和噪声,用回归模型比较两种算法的影响。
关键影像表现/征象
- 近植入物处,iMAR较常规ADMIRE在骨、肌肉、脂肪的CT值上均有显著差异(P<0.001)。
- 近植入物处,iMAR较常规ADMIRE在骨噪声上有显著差异(P<0.003)。
- 远植入物处,iMAR与ADMIRE在骨、肌肉、脂肪三种组织的CT值和噪声均无显著差异。
临床应用与阅片要点
- 儿童带骨科金属植入物的低剂量CT重建可关注iMAR算法对近植入物条纹伪影的改善作用。
- 评估iMAR效果时宜对照远植入物处的组织CT值和噪声,确认其基线组织密度未改变。
局限性
- 原文结论称上述为初步结果(preliminary results)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Increased computational power allows computed tomography (CT) software to process very advanced mathematical algorithms to generate better quality images at lower doses. One such algorithm, iterative metal artifact reduction (iMAR) has proven to decrease metal artifacts seen in CT images of adults with orthopedic implants.
OBJECTIVES: To evaluate artifact reduction capability of the algorithm in lower-dose pediatric CT compared to our routine third-generation advanced modeled iterative reconstruction (ADMIRE) algorithm.
MATERIALS AND METHODS: Thirteen children (11-17 years old) with metal implants underwent routine clinically indicated CT. Data sets were reconstructed with an iMAR algorithm. Hounsfield units and image noise were measured in bone, muscle and fat in the streak artifact (near the implant) and at the greatest distance from the artifact (far from the implant). A regression model compared the effects of the algorithm (standard ADMIRE vs. iMAR) near and far from the implant.
RESULTS: Near the implant, Hounsfield units with iMAR were significantly different in our standard ADMIRE vs. iMAR for bone, muscle and fat (P<0.001). Noise was significantly different in standard ADMIRE vs. iMAR in bone (P<0.003). Far from the implant, Hounsfield units and noise were not significantly different for ADMIRE vs. iMAR, for the three tissue types.
CONCLUSION: These preliminary results demonstrate that iMAR algorithms improves Hounsfield units near the implant and decreases image noise in bone in low-dose pediatric CT. It does this without changing baseline tissue density or noise far from the implant.