基于图像模拟的低剂量计算机断层扫描图像提示13 mAs 120 kV适用于非综合征性颅缝早闭诊断且无需迭代重建算法
本研究纳入29例因颅骨畸形行头部CT的儿童患者,通过向原始数据添加噪声合成120 kV下120、100、80、50及13 mAs的低剂量图像,由3名研究者用问卷评估诊断适用性。结果显示13 mAs模拟图像诊断适用率为96%,与原始图像98.9%相近;结论认为120 kV 13 mAs图像诊断非综合征性颅缝早闭的准确性与标准方案统计学相同,有效剂量中位数从4.98 mSv降至0.33 mSv。
要点
- 120 kV 13 mAs模拟图像诊断适用率96%,原始图像为98.9%。
- 13 mAs方案有效剂量中位数0.33 mSv,标准方案为4.98 mSv。
- 低至13 mAs的模拟图像无需迭代重建算法即可用于诊断。
核心结论
- 120 kV 13 mAs采集的图像可用于诊断非综合征性颅缝早闭,其准确性与标准方案统计学相同,有效剂量中位数从4.98 mSv降至0.33 mSv。
研究设计与样本
- 研究纳入29例因颅骨畸形行头部CT检查的儿童患者,通过向原始数据添加噪声合成低剂量图像。
- 合成设置对应120 kV下120 mAs、100 mAs、80 mAs、50 mAs和13 mAs,由3名研究者用问卷评估诊断适用性。
关键影像表现/征象
- 共评估174份独立病例(含1幅轴位和1幅3D图像)。
- 评估为适合诊断的图像比例:原始图像98.9%,120 mAs为100%,100 mAs为100%,80 mAs为97.1%,50 mAs为96.6%,13 mAs为96%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to simulate low dose paediatric head CT images with different noise levels corresponding to various tube current time product values and assess simulated image suitability in non-syndromic craniosynostosis diagnostics.
METHOD: 29 paediatric patients who underwent head CT examinations for cranial deformity were enrolled in the study. The low dose CT images, corresponding to 120 kV and 120 mAs, 100 mAs, 80 mAs, 50 mAs and 13 mAs settings, were synthesised by adding noise to original data. Three researchers evaluated suitability for diagnostics of original and simulated images by using questionnaire assessing image suitability.
RESULT: 174 separate cases (containing 1 axial and 1 3D image) were evaluated. Percentage of images evaluated as suitable for diagnosis were 98.9% on original images, 100% on 120 mAs, 100% on 100 mAs, 97.1% on 80 mAs, 96.6% on 50 mAs and 96% on 13 mAs.
CONCLUSIONS: Images registered with 120 kV 13 mAs can be used to diagnose non-syndromic craniosynostosis with statistically same accuracy as with standard protocol and correspond to decrease of effective dose from 4.98 mSv to 0.33 mSv (median values).
引用
- Neverauskiene A, Maciusovic M, Burkanas M, et al. Image based simulation of the low dose computed tomography images suggests 13 mAs 120 kV suitability for non-syndromic craniosynostosis diagnosis without iterative reconstruction algorithms. Eur J Radiol. 2018. PMID: 30017275. doi:10.1016/j.ejrad.2018.06.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。