健康志愿者正常肱骨外上髁伸肌腱的剪切波弹性成像观察者间变异性与硬度测量

Interobserver variability and stiffness measurements of normal common extensor tendon in healthy volunteers using shear wave elastography.
PubMed 文献 · Skeletal Radiol · 2018-07-13

本研究对30名健康志愿者(15女15男,平均年龄30.2岁)共60条无肱骨外上髁炎症状的肱骨外上髁伸肌腱行剪切波弹性成像,由两名放射科医生测量硬度。正常肌腱平均厚度3.57±0.36 mm,两名观察者平均硬度分别为45.28±9.82 kPa和45.80±9.72 kPa,观察者间测量误差范围为±23.5%,绝对测量值间无显著差异(p=0.741)。

要点

  • 正常肱骨外上髁伸肌腱平均硬度约45 kPa,可作为鉴别正常与病理组织的参考值。
  • 肌腱厚度与硬度呈中等相关(r=0.429,p<0.001),与体重呈弱相关(r=0.281,p=0.03)。
  • 观察者间测量误差范围为±23.5%,绝对测量值间无显著差异(p=0.741)。

核心结论

  • 剪切波弹性成像评估肱骨外上髁伸肌腱弹性具有可重复性,正常肌腱的标准硬度值可作为鉴别正常与病理组织的参考数据。

研究设计与样本

  • 对30名健康志愿者(15女15男,平均年龄30.2岁)共60条无肱骨外上髁炎症状的肱骨外上髁伸肌腱行B模式和剪切波弹性成像,由两名放射科医生测量,第一观察者行B模式和剪切波弹性成像,第二观察者仅行剪切波弹性成像。

关键影像表现/征象

  • 正常肱骨外上髁伸肌腱平均厚度为3.57±0.36 mm。
  • 两名观察者测得的正常肱骨外上髁伸肌腱平均硬度分别为45.28±9.82 kPa和45.80±9.72 kPa。

诊断标准与关键数值

  • 正常肱骨外上髁伸肌腱硬度参考值:第一观察者45.28±9.82 kPa,第二观察者45.80±9.72 kPa。
  • 观察者间硬度测量平均差异为平均值的-0.5%,测量误差范围(95%一致性界限)为±23.5%。
  • 观察者间绝对测量值间无显著差异(p=0.741)。

临床应用与阅片要点

  • 剪切波弹性成像可用于评估肱骨外上髁伸肌腱弹性,正常肌腱标准硬度值可作为鉴别正常与病理组织的参考数据。
  • 观察者间测量误差范围为±23.5%,判读硬度值时需考虑该测量变异。
  • 肌腱厚度与硬度值呈中等相关(r=0.429,p<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purposes of our study are to determine the quantitative elasticity values of normal common extensor tendon (CET) and to assess the interobserver variability of stiffness measurements using shear wave elastography (SWE).

MATERIALS AND METHODS: A total of 60 CETs of 30 (15 female, 15 male, mean age 30.2 years) healthy volunteers without any symptoms of lateral epicondylitis were examined by two radiologists. Age, sex, height, weight, body mass index (BMI), and dominant hand of all participants were noted. The first observer performed B-mode and SWE imaging, and the second observer performed only SWE imaging. Tendon thickness and stiffness values in kPa were measured.

RESULTS: The mean thickness of CETs was 3.57 ± 0.36 mm. The mean stiffness values of CETs for two observers were 45.28 ± 9.82 kPa and 45.80 ± 9.72 kPa respectively. Tendon thickness had a weak correlation with weight (r = 0.281, p = 0.03), and moderate correlation with stiffness values (r = 0.429, p < 0.001). The mean interobserver difference of CET stiffness measurements was -0.5% of the mean CET stiffness values. Range of measurement error, defined as 95% limits of agreement, was ±23.5%. There was no significant difference between absolute values of interobserver measurements (p = 0.741).

CONCLUSION: Shear wave elastography is a reproducible imaging technique for the evaluation of CET elasticity and the standard stiffness values of normal CET can be used as reference data to differentiate normal from pathological tissues.

来源:Skeletal Radiol,查看原文。版权归原作者所有。