光声成像对良恶性乳腺肿块灰阶超声BI-RADS分类的降级与升级:一项初步研究
这项前瞻性多中心初步研究纳入94个乳腺肿块,评估光声超声相较于内部灰阶超声对BI-RADS分类的增量价值;光声超声与灰阶超声敏感度均为97.1%,特异度分别为44.3%和36.4%;光声超声使41.7%的良性或稳定BI-RADS 3类肿块降级为2类,并使75.0%的4A至5类恶性肿块升级。
要点
- 光声超声与灰阶超声敏感度相同(97.1%),特异度分别为44.3%和36.4%。
- 光声超声使41.7%的良性或稳定BI-RADS 3类肿块降级为BI-RADS 2类。
- 光声超声使75.0%的4A至5类恶性肿块升级,49.4%的恶性肿块归为4B至5类。
核心结论
- 与灰阶超声相比,光声超声可使良性乳腺肿块BI-RADS分类降级、恶性肿块分类升级。
研究设计与样本
- 前瞻性多中心初步研究,纳入94个乳腺肿块(39个活检证实恶性、44个活检证实良性、11个BI-RADS 3类且12个月随访稳定),设备为手持双模激光与内置灰阶超声探头的光声乳腺成像装置,由独立阅片放射科医生评估内部灰阶超声和光声超声特征并赋予BI-RADS分类。
诊断标准与关键数值
- 光声超声与内部灰阶超声的敏感度均为97.1%;特异度光声超声为44.3%,内部灰阶超声为36.4%;光声超声使41.7%的良性或12个月随访稳定的BI-RADS 3类肿块降级为BI-RADS 2类;36.6%的BI-RADS 4A类肿块降级为3类或2类;10.1%的BI-RADS 4B类肿块降级为3类或2类;独立阅片者用光声超声将75.0%的4A、4B、4C或5类恶性肿块升级;49.4%的恶性肿块归为4B、4C或5类。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: False-positive findings remain challenging in breast imaging. This study investigates the incremental value of optoacoustic imaging in improving BI-RADS categorization of breast masses at ultrasound.
SUBJECTS AND METHODS: The study device is an optoacoustic breast imaging device with a handheld duplex laser and internal gray-scale ultrasound probe, fusing functional and morphologic information (optoacoustic ultrasound). In this prospective multisite study, breast masses assessed as BI-RADS category 3, 4A, 4B, 4C, or 5 by site radiologists underwent both gray-scale ultrasound and optoacoustic imaging with the study device. Independent reader radiologists assessed internal gray-scale ultrasound and optoacoustic ultrasound features for each mass and assigned a BI-RADS category. The percentage of mass reads for which optoacoustic ultrasound resulted in a downgrade or upgrade of BI-RADS category relative to internal gray-scale ultrasound was determined.
RESULTS: Of 94 total masses, 39 were biopsy-proven malignant, 44 were biopsy-proven benign, and 11 BI-RADS category 3 masses were stable at 12-month follow-up. The sensitivity of both optoacoustic ultrasound and internal gray-scale ultrasound was 97.1%. The specificity was 44.3% for optoacoustic ultrasound and 36.4% for internal gray-scale ultrasound. Using optoacoustic ultrasound, 41.7% of benign masses or BI-RADS category 3 masses that were stable at 12-month follow-up were downgraded to BI-RADS category 2 by independent readers; 36.6% of masses assigned BI-RADS category 4A were downgraded to BI-RADS category 3 or 2, and 10.1% assigned BI-RADS category 4B were downgraded to BI-RADS category 3 or 2. Using optoacoustic ultrasound, independent readers upgraded 75.0% of the malignant masses classified as category 4A, 4B, 4C, or 5, and 49.4% of the malignant masses were classified as category 4B, 4C, or 5.
CONCLUSION: Optoacoustic ultrasound resulted in BI-RADS category downgrading of benign masses and upgrading of malignant masses compared with gray-scale ultrasound.