本研究回顾性分析39例患者共150幅传统力度标准化压迫乳腺X线摄影(FSCM)图像与150幅压力标准化压迫乳腺X线摄影(PSCM)图像进行个体内比较,PSCM使平均压力从17.1 kPa显著降至12.8 kPa(p < 0.001),过度压迫相对数量从26%降至2%,提示PSCM可减轻不适与疼痛并可能提供更稳定的图像质量。
要点
- PSCM将平均压迫压力从FSCM的17.1 kPa降至12.8 kPa(p < 0.001)。
- 过度压迫相对数量从FSCM的26%降至PSCM的2%,小乳房患者获益。
- PSCM下压力几乎不依赖乳房接触面积(-0.014 kPa/cm2),FSCM则明显依赖。
核心结论
- 研究提示,与常规FSCM相比,使用PSCM可因平均压力降低而减少乳腺摄影压迫过程中的患者不适和疼痛,标准化压力可能提供更恒定的图像质量,从而可能改善诊断性能。
研究设计与样本
- 回顾性研究,个体内比较39例患者,每例患者双乳各接受一次头尾位(CC)和一次内外侧斜位(MLO)压迫,分别使用传统力度标准化压迫乳腺X线摄影(FSCM)技术和压力标准化压迫乳腺X线摄影(PSCM)技术,共分析150幅FSCM和150幅PSCM图像,所有图像处理后获得乳房与压迫板之间的接触面积,压力通过压迫力除以接触面积计算。
诊断标准与关键数值
- 平均压力从FSCM的17.1 kPa降至PSCM的12.8 kPa(p < 0.001)。
- PSCM下压力与乳房接触面积的相关性为-0.014 kPa/cm2,FSCM下观察到明显依赖。
- 过度压迫相对数量从FSCM的26%降至PSCM的2%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Validation of a pressure-standardized compression mammography (PSCM) system, which aims to reduce discomfort and pain by applying the same pressure to every breast, independent of breast size.
METHODS: We retrospectively studied mammograms of 39 patients acquired with a conventional force-standardized compression mammography (FSCM) technique and intra-individually compared them to mammograms acquired on a checkup visit with PSCM technique. Patients received one craniocaudal (CC) and one mediolateral oblique (MLO) compression for both breasts. All images were processed to obtain the contact area between the breast and the compression paddle. The pressure was calculated by dividing the compression force by the contact area.
RESULTS: A total of 150 FSCM and 150 PSCM images were analyzed. The mean pressure decreased significantly from 17.1 to 12.8 kPa (p < 0.001), when using PSCM instead of FSCM. The applied pressure hardly depended on the breast contact area with the paddle (-0.014 kPa/cm2), while a clear dependency was observed using FSCM. Furthermore, the relative number of over-compressions reduced from 26% to 2%, benefitting patients with smaller breasts.
CONCLUSIONS: Our study suggests that using PSCM can reduce patient discomfort and pain during mammographic compression compared to conventional FSCM as a result of lower average pressure. Moreover, standardized pressure may provide a more constant image quality, which could improve diagnostic performance.
引用
- den Boer D, Dam-Vervloet LAJ, Boomsma MF, et al. Clinical validation of a pressure-standardized compression mammography system. Eur J Radiol. 2018. PMID: 30017290. doi:10.1016/j.ejrad.2018.06.021