基波与二次谐波叠加结合百微秒超声脉冲的组织摧毁术
本研究提出一种基于基波与二次谐波叠加、采用百微秒级长脉冲和两阶段脉冲方案的组织摧毁术新方法,在含牛血清白蛋白的聚丙烯酰胺凝胶体模和离体猪组织中验证可行性;离体猪肾中可形成轴向5.80±0.19 mm、横向2.20±0.26 mm的规则长撕裂形完全组织匀质化区域,组织学证实病灶彻底匀质化且与未处理区边界清晰。
要点
- 采用1.1/2.2 MHz双元件共焦环形阵列,焦域峰压分别为+22/-7 MPa和+56/-14 MPa(含冲击波)。
- 两阶段脉冲方案可快速生成机械匀质化病灶,离体猪肾病灶轴向5.80±0.19 mm、横向2.20±0.26 mm。
- HE染色显示病灶彻底匀质化且与未处理区边界清晰,作者认为该方法可能成为精确肿瘤治疗工具。
核心结论
- 基波与二次谐波叠加结合百微秒脉冲和两阶段脉冲方案的组织摧毁术,能以空间精确性高效实现软组织机械性破坏,该方法可能有潜力发展为精确肿瘤治疗的有用工具。
研究设计与样本
- 可行性实验研究,在含牛血清白蛋白的聚丙烯酰胺凝胶体模和离体猪组织中进行;使用定制1.1/2.2 MHz双元件共焦环形阵列,f数为0.69,半高全宽压力尺寸横向约1.0 mm、轴向约6.0 mm;通过高速摄影观察空化和沸腾活动,被动空化检测记录声发射。
诊断标准与关键数值
- 两阶段脉冲方案:阶段1脉冲持续时间500-600 μs、脉冲重复频率100 Hz;阶段2由多个周期组成,每周期含多个与阶段1相同脉冲持续时间和重复频率的脉冲,周期之间关闭时间600 ms。
- 焦域峰正/负压:1.1 MHz为+22/-7 MPa;2.2 MHz含冲击波为+56/-14 MPa。
- 离体猪肾完全组织匀质化区域呈规则长撕裂形,典型尺寸轴向5.80±0.19 mm、横向2.20±0.26 mm。
局限性
- 本研究为可行性实验,仅在组织仿体模型和离体猪组织中验证,作者结论为该方法可能有潜力发展为精确肿瘤治疗工具。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
A novel histotripsy approach based on fundamental and second harmonic superposition and incorporating hundred-microsecond-long pulses and two-stage pulse protocol is proposed in this study to rapidly generate mechanically homogenized lesions. Two pulse stages were applied: stage 1, pulses with a pulse duration of 500-600 μs and pulse repetition frequency of 100 Hz, and stage 2, multiple periods, each composed of multiple pulses with the same pulse duration and pulse repetition frequency as those in stage 1, but with an off-time of 600 ms between periods. A custom-designed 1.1/2.2-MHz two-element confocal-annular array, with an f-number of 0.69, and lateral and axial full width at half-maximum pressure dimensions of approximately 1.0 and 6.0 mm, was used. The peak positive/negative pressures at the focus were +22/-7 MPa for 1.1 MHz and +56/-14 MPa with shock wave for 2.2 MHz. To investigate the feasibility of this approach, experiments were designed and performed in tissue-mimicking polyacrylamide gel phantoms with bovine serum albumin and in ex vivo porcine tissues. Cavitation and boiling activities were observed through high-speed photography, and the corresponding acoustic emissions were recorded through passive cavitation detection. Ex vivo experimental results revealed that complete tissue homogeneous regions with regular long tear shape and typical dimensions of 5.80 ± 0.19 mm in axial and 2.20 ± 0.26 mm in lateral were successfully generated in porcine kidney samples. The hematoxylin and eosin staining evidenced that the lesions were thoroughly homogenized and sharply demarcated from untreated regions. These results indicated that the histotripsy approach using fundamental and second harmonic superposition combined with hundred-microsecond pulses and two-stage pulse protocol can efficiently obtain a mechanical disruption of soft tissues with spatial precision, and this approach may have the potential to be developed as a useful tool for precise tumor treatment.
引用
- Li Y, Wang R, Lu M, et al. Histotripsy Using Fundamental and Second Harmonic Superposition Combined with Hundred-Microsecond Ultrasound Pulses. Ultrasound Med Biol. 2018. PMID: 30054023. doi:10.1016/j.ultrasmedbio.2018.05.024
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。