文献速递腹部技术与安全US

慢性乙型肝炎患者肝脏纤维化的二维超声剪切波弹性成像:一项前瞻性多中心研究

Liver Fibrosis with Two-dimensional US Shear-Wave Elastography in Participants with Chronic Hepatitis B: A Prospective Multicenter Study.
PubMed 文献 · Radiology · 2018-07-24

这项前瞻性多中心研究纳入402例慢性乙型肝炎参与者,比较二维剪切波弹性成像(2D SWE)与瞬时弹性成像(TE)、APRI及FIB-4对肝硬化的诊断性能。2D SWE诊断肝硬化的AUC为0.87,高于TE的0.80、APRI的0.70和FIB-4的0.73;在慢性感染组AUC达0.92,显著高于慢性肝炎组的0.84(P=.017)。

要点

  • 2D SWE诊断肝硬化的AUC为0.87,高于TE的0.80、APRI的0.70和FIB-4的0.73。
  • 慢性感染组肝硬化诊断AUC为0.92,显著高于慢性肝炎组的0.84(P=.017)。
  • 慢性感染组以8.4 kPa和11.0 kPa双截断值排除和确认肝硬化,81.2%(125/154)可明确诊断。

核心结论

  • 2D SWE的诊断性能高于其他无创方法,在慢性感染参与者中排除和确认肝硬化最为有效,可能促进抗病毒治疗。

研究设计与样本

  • 前瞻性多中心研究,2015年1月至2016年1月在12家医院开展,纳入654例接受肝活检和2D SWE检查的慢性乙型肝炎参与者,最终402例入组(慢性感染154例、慢性肝炎248例)。
  • 以肝活检为参考标准,采用DeLong检验比较2D SWE与APRI、FIB-4及TE的诊断性能,并用多级似然比分析确定肝硬化的双截断值。

诊断标准与关键数值

  • 2D SWE诊断肝硬化的AUC为0.87(95%CI:0.83,0.90),高于TE的0.80(95%CI:0.68,0.88)、APRI的0.70(95%CI:0.65,0.74)和FIB-4的0.73(95%CI:0.69,0.78)。
  • 慢性感染组诊断肝硬化的AUC为0.92(95%CI:0.87,0.96),显著高于慢性肝炎组的0.84(95%CI:0.78,0.88;P=.017)。
  • 慢性感染组确定双截断值8.4 kPa和11.0 kPa分别用于排除和确认肝硬化,对应似然比低于0.1和高于10,81.2%(125/154)的肝硬化参与者可明确诊断。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To investigate the diagnostic performance of two-dimensional (2D) shear-wave elastography (SWE) in chronic hepatitis B. Materials and Methods This prospective multicenter study from January 2015 to January 2016 was conducted at 12 hospitals and included 654 participants with chronic hepatitis B who had undergone liver biopsy and 2D SWE examination. Participants were divided into chronic infection and chronic hepatitis groups. The diagnostic performance of 2D SWE was compared with the aspartate amino transferase-to-platelet ratio index (APRI), the Fibrosis-4 index (FIB-4), and transient elastography (TE) by using a DeLong test and was also compared between two subgroups. Dual cutoff values for cirrhosis were determined with multilevel likelihood ratio analysis. Results Overall, 402 participants with chronic hepatitis B were enrolled (154 with chronic infection and 248 with chronic hepatitis). The areas under the receiver operating characteristic curve of 2D SWE (0.87; 95% confidence interval [CI]: 0.83, 0.90) were higher than those of TE (0.80; 95% CI: 0.68, 0.88), APRI (0.70; 95% CI: 0.65, 0.74), and FIB-4 (0.73; 95% CI: 0.69, 0.78) in cirrhosis. The high area under the receiver operating characteristic curve (0.92; 95% CI: 0.87, 0.96) was achieved in the chronic infection group and was significantly higher than that of the chronic hepatitis group (0.84; 95% CI: 0.78, 0.88; P = .017). Dual cutoff values with the likelihood ratios below 0.1 and above 10 (8.4 kPa and 11.0 kPa to rule out and rule in a diagnosis of cirrhosis, respectively) were effectively determined in chronic infection; a total of 81.2% (125 of 154) participants with cirrhosis were definitively diagnosed. Conclusion The performance of two-dimensional (2D) shear-wave elastography (SWE) was higher than that of other noninvasive methods. 2D SWE was most effective in ruling in and ruling out cirrhosis in participants with chronic infection, which may prompt antiviral treatment. © RSNA, 2018 Online supplemental material is available for this article.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Gao Y, Zheng J, Liang P, et al. Liver Fibrosis with Two-dimensional US Shear-Wave Elastography in Participants with Chronic Hepatitis B: A Prospective Multicenter Study. Radiology. 2018. PMID: 30040048. doi:10.1148/radiol.2018172479

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。