二维剪切波弹性成像测量脾脏硬度:测量成功率的预测因素与正常脾脏硬度值
本研究对313例连续患者用二维剪切波弹性成像测量脾脏和肝脏硬度,总体测量成功率为52.9%,腹壁厚度和脾脏长径是测量成功的独立预测因素,正常患者组平均脾脏硬度为20.5±5.4 kPa。
要点
- 脾脏硬度测量总体成功率为52.9%,低于肝脏硬度测量成功率。
- 腹壁厚度≤17.2 mm和脾脏长径>9.4 cm是测量成功的独立预测因素。
- 正常患者组平均脾脏硬度值为20.5±5.4 kPa。
核心结论
- 脾脏硬度测量成功率较低,受腹壁厚度和脾脏大小影响,正常脾脏硬度平均值为20.5±5.4 kPa。
研究设计与样本
- 研究纳入313例连续患者,采用二维剪切波弹性成像测量脾脏和肝脏硬度,并评估体重指数、腹壁厚度、脾脏大小和肝脏硬度与测量成功的关系。
诊断标准与关键数值
- 测量成功的独立预测因素为腹壁厚度≤17.2 mm和脾脏长径>9.4 cm;正常患者组平均脾脏硬度值为20.5±5.4 kPa。
临床应用与阅片要点
- 进行脾脏硬度测量时,需注意腹壁厚度和脾脏大小对测量成功率的影响。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the success rate of spleen stiffness measurement and factors that affect measurement success and to determine the spleen stiffness value of normal individuals by 2-dimensional shear wave elastography.
METHODS: The spleen and liver stiffnesses of 313 consecutive patients were measured with SWE. The body mass index, abdominal wall thickness, spleen size, and liver stiffness of the patients were evaluated to identify factors associated with successful measurement of spleen stiffness. Patients were grouped by body mass index, spleen size, and liver stiffness, and the success rates and mean spleen stiffness values of the groups were compared. Independent predictors for successful spleen stiffness measurement and their cutoff values were evaluated. The mean spleen stiffness values of patients considered to have normal spleen stiffness were investigated.
RESULTS: The overall success rate of spleen stiffness measurement was 52.9%. It was significantly higher in nonobese than in obese patients and in patients with splenomegaly and liver cirrhosis. The spleen stiffness value was higher in the splenomegaly group than the nonsplenomegaly group (P < .001) and increased as liver stiffness increased (P < .001). There was no significant difference in spleen stiffness values between the obese and nonobese groups. Abdominal wall thickness and splenic longitudinal diameter were identified as independent predictors of successful spleen stiffness measurement, and their cutoff values were 17.2 mm or less and greater than 9.4 cm, respectively. The mean spleen stiffness value ± SD of the normal patient group was 20.5 ± 5.4 kPa.
CONCLUSIONS: The success rate of spleen stiffness measurement is lower than that of liver stiffness measurement. Spleen stiffness measurement is affected by abdominal wall thickness and spleen size.